Reproducibility of cardiac output measurement by cross sectional and Doppler echocardiography
- PMID: 3395526
- PMCID: PMC1276875
- DOI: 10.1136/hrt.59.6.680
Reproducibility of cardiac output measurement by cross sectional and Doppler echocardiography
Abstract
The variability of Doppler echocardiographic estimation of cardiac output at the aortic orifice was investigated in eight healthy subjects. Cross sectional echocardiograms of the aortic orifice and aortic Doppler velocities were recorded and measured by four echocardiographers. Between subject variability was significantly larger than within subject variability for all variables. Variability owing to different echocardiographers and different measurement times was small compared with total variability. Coefficients of variation for aortic annular diameter, aortic velocity integral, and heart rate were 4.1%, 6.4%, and 5.0% respectively. The coefficient of variation for cardiac output was 8.8% and the 95% confidence interval for measurement of cardiac output by the Doppler method was 4.45 to 6.35 l/min. One echocardiographer reanalysed all the recordings and the results showed that recording the echocardiograms introduces a significantly larger source of error than measuring them. Thus serial measurements of cardiac output by the Doppler method can be performed with acceptable reproducibility; this indicates that the technique can be accurately applied in clinical practice.
Similar articles
-
Doppler echocardiographic estimation of cardiac output: analysis of temporal variability.Eur Heart J. 1988 Mar;9(3):313-8. doi: 10.1093/oxfordjournals.eurheartj.a062502. Eur Heart J. 1988. PMID: 3383871
-
Cardiac output determinations in the newborn. Reproducibility of the pulsed Doppler velocity measurement.J Ultrasound Med. 1988 Jun;7(6):311-5. doi: 10.7863/jum.1988.7.6.311. J Ultrasound Med. 1988. PMID: 2969423
-
Measurement of left ventricular stroke volume using continuous wave Doppler echocardiography of the ascending aorta and M-mode echocardiography of the aortic valve.J Am Coll Cardiol. 1987 Jan;9(1):75-83. doi: 10.1016/s0735-1097(87)80085-2. J Am Coll Cardiol. 1987. PMID: 3794113
-
Doppler echocardiography for the estimation of cardiac output with exercise.Sports Med. 2002;32(15):973-86. doi: 10.2165/00007256-200232150-00002. Sports Med. 2002. PMID: 12457418 Review.
-
[Noninvasive measurement of cardiac output by pulsed Doppler echocardiography. Correlation with thermodilution].Arch Mal Coeur Vaiss. 1990 Feb;83(2):237-44. Arch Mal Coeur Vaiss. 1990. PMID: 2106860 Review. French.
Cited by
-
Precision of Echocardiographic Measurements.J Am Soc Echocardiogr. 2024 May;37(5):562-563. doi: 10.1016/j.echo.2024.01.001. Epub 2024 Jan 8. J Am Soc Echocardiogr. 2024. PMID: 38199333 Free PMC article.
-
Three dimensional colour Doppler echocardiography for the characterisation and quantification of cardiac flow events.Heart. 2000 Nov;84 Suppl 2(Suppl 2):II2-6. doi: 10.1136/heart.84.suppl_2.ii2. Heart. 2000. PMID: 11040028 Free PMC article. Review. No abstract available.
-
Non-invasive automated assessment of the ratio of pulmonary to systemic flow in patients with atrial septal defects by the colour Doppler velocity profile integration method.Heart. 2002 Sep;88(3):278-82. doi: 10.1136/heart.88.3.278. Heart. 2002. PMID: 12181223 Free PMC article.
-
Comparison of the relative airways and systemic potencies of inhaled fenoterol and salbutamol in asthmatic patients.Thorax. 1995 Jan;50(1):54-61. doi: 10.1136/thx.50.1.54. Thorax. 1995. PMID: 7886650 Free PMC article. Clinical Trial.
-
Cardiac index measurements by transcutaneous Doppler ultrasound and transthoracic echocardiography in adult and pediatric emergency patients.J Clin Monit Comput. 2010 Jun;24(3):237-47. doi: 10.1007/s10877-010-9240-6. Epub 2010 Jun 20. J Clin Monit Comput. 2010. PMID: 20563629
References