A new dynamometer measuring concentric and eccentric muscle strength in accelerated, decelerated, or isokinetic movements. Validity and reproducibility
- PMID: 3396567
- DOI: 10.1007/BF00418457
A new dynamometer measuring concentric and eccentric muscle strength in accelerated, decelerated, or isokinetic movements. Validity and reproducibility
Abstract
A new computerized dynamometer (the SPARK System) is described. The system can measure concentric and eccentric muscle strength (torque) during linear or nonlinear acceleration or deceleration, isokinetic movements up to 400 degrees.s-1, and isometric torque. Studies were performed to assess: I. validity and reproducibility of torque measurements; II. control of lever arm position; III. control of different velocity patterns; IV. control of velocity during subject testing; and, V. intra-individual reproducibility. No significant difference was found between torque values computed by the system and known torque values (p greater than 0.05). No difference was present between programmed and external measurement of the lever arm position. Accelerating, decelerating and isokinetic velocity patterns were highly reproducible, with differences in elapsed time among 10 trials being never greater than 0.001 s. Velocity during concentric and eccentric isokinetic quadriceps contractions at 30 degrees.s-1, 120 degrees.s-1 and 270 degrees.s-1 never varied by more than 3 degrees.s-1 among subjects (N = 21). Over three days of testing, the overall error for concentric and eccentric quadriceps contraction peak torque values for 5 angular velocities between 30 degrees.s-1 and 270 degrees.s-1 ranged from 5.8% to 9.0% and 5.8% to 9.6% respectively (N = 25). The results indicate that the SPARK System provides valid and reproducible torque measurements and strict control of velocity. In addition, the intra-individual error is in accordance with those reported for other similar devices.
Similar articles
-
Eccentric and concentric torque-velocity characteristics of the quadriceps femoris in man.Eur J Appl Physiol Occup Physiol. 1988;58(1-2):100-4. doi: 10.1007/BF00636611. Eur J Appl Physiol Occup Physiol. 1988. PMID: 3203653
-
Differences in elbow flexion torque measured concentrically, eccentrically, and isometrically.Phys Ther. 1987 Aug;67(8):1205-8. doi: 10.1093/ptj/67.8.1205. Phys Ther. 1987. PMID: 3615588
-
Continuous versus separate isokinetic test protocol: the effect of estradiol on the reproducibility of concentric and eccentric isokinetic measurements in knee muscles.Arch Phys Med Rehabil. 1999 Sep;80(9):1024-9. doi: 10.1016/s0003-9993(99)90055-2. Arch Phys Med Rehabil. 1999. PMID: 10489003 Clinical Trial.
-
Isokinetic dynamometry: implications for muscle testing and rehabilitation.Exerc Sport Sci Rev. 1986;14:45-80. Exerc Sport Sci Rev. 1986. PMID: 3525192 Review.
-
Hamstrings to quadriceps peak torque ratios diverge between sexes with increasing isokinetic angular velocity.J Sci Med Sport. 2008 Sep;11(5):452-9. doi: 10.1016/j.jsams.2007.04.009. Epub 2007 Sep 17. J Sci Med Sport. 2008. PMID: 17875402 Free PMC article. Review.
Cited by
-
Strength and power assessment. Issues, controversies and challenges.Sports Med. 1995 Jun;19(6):401-17. doi: 10.2165/00007256-199519060-00004. Sports Med. 1995. PMID: 7676101 Review.
-
Static and dynamic assessment of the Biodex dynamometer.Eur J Appl Physiol Occup Physiol. 1991;62(3):180-8. doi: 10.1007/BF00643739. Eur J Appl Physiol Occup Physiol. 1991. PMID: 2044524
-
Effects of prolonged exercise on the contractile properties of human quadriceps muscle.Eur J Appl Physiol Occup Physiol. 1995;71(2-3):180-6. doi: 10.1007/BF00854977. Eur J Appl Physiol Occup Physiol. 1995. PMID: 7588687
-
Responsiveness of Isokinetic Dynamometry in Patients with Osteoarthritis after Knee and Hip Arthroplasty: A Prospective Repeated-Measures Cohort Study.Healthcare (Basel). 2024 Jan 25;12(3):314. doi: 10.3390/healthcare12030314. Healthcare (Basel). 2024. PMID: 38338199 Free PMC article.
-
The utility of isokinetic dynamometry in the assessment of human muscle function.Sports Med. 1996 Jan;21(1):18-34. doi: 10.2165/00007256-199621010-00003. Sports Med. 1996. PMID: 8771283 Review.