Validation of an EEG seizure detection paradigm optimized for clinical use in a chronically implanted subcutaneous device
- PMID: 33971201
- DOI: 10.1016/j.jneumeth.2021.109220
Validation of an EEG seizure detection paradigm optimized for clinical use in a chronically implanted subcutaneous device
Abstract
Background: Many electroencephalography (EEG) based seizure detection paradigms have been developed and validated over the last two decades. The majority of clinical approaches use scalp or intracranial EEG electrodes. Scalp EEG is limited by patient discomfort and short duration of useful EEG signals. Intracranial EEG involves an invasive surgical procedure associated with significant risk making it unsuitable for widespread use as a practical clinical biometric. A less invasive EEG monitoring approach, that is between invasive intracranial procedures and noninvasive methods, would fill the need of a safe, accurate, chronic (ultra-long term) and objective seizure detection method. We present validation of a continuous EEG seizure detection paradigm using human single-channel EEG recordings from subcutaneously placed electrodes that could be used to fulfill this need.
Methods: Ten-minute long sleep, awake and ictal EEG epochs obtained from 21 human subjects with subscalp electrodes and validated against simultaneous iEEG recordings were analyzed by three experienced clinical neurophysiologists. The 201subscalp EEG time series epochs where classified as diagnostic for awake, asleep, or seizure by the clinicians who were blinded to all other information. Seventy of the epochs were classified in this way as representing seizure activity. A subject specific seizure detection algorithm was trained and then evaluated offline for each patient in the data set using the expert consensus classification as the gold standard.
Results: The average seizure detection performance of the algorithm across 21 subjects exceeded 90 % accuracy: 97 % sensitivity, 91 % specificity, and 93 % accuracy. For 19 of 21 patient datasets the algorithm achieved 100 % sensitivity. For 15 of 21 patients, the algorithm achieved 100 % specificity. For 13 of 21 patients the algorithm achieved 100 % accuracy.
Comparison: No comparable published methods are available for subgaleal EEG seizure detection.
Conclusions: These findings suggest that a simple seizure detection algorithm using subcutaneous EEG signals could provide sufficient accuracy and clinical utility for use in a low power, long-term subcutaneous brain monitoring device. Such a device would fill a need for a large number of people with epilepsy who currently have no means for accurately quantifying their seizures thereby providing important information to healthcare providers not currently available.
Keywords: Algorithm; Band-pass filters; EEG; Electroencephalography; Epilepsy; Intracranial; Linear detector; Low power; Scalp; Subgaleal; Ultra-long-term EEG monitoring.
Copyright © 2021 Elsevier B.V. All rights reserved.
Similar articles
-
Intracranial EEG Validation of Single-Channel Subgaleal EEG for Seizure Identification.J Clin Neurophysiol. 2022 May 1;39(4):283-288. doi: 10.1097/WNP.0000000000000774. Epub 2020 Nov 20. J Clin Neurophysiol. 2022. PMID: 32925251
-
Seizure detection with deep neural networks for review of two-channel electroencephalogram.Epilepsia. 2023 Dec;64 Suppl 4:S34-S39. doi: 10.1111/epi.17259. Epub 2022 Apr 24. Epilepsia. 2023. PMID: 35416283
-
Electromyography-based seizure detector: Preliminary results comparing a generalized tonic-clonic seizure detection algorithm to video-EEG recordings.Epilepsia. 2015 Sep;56(9):1432-7. doi: 10.1111/epi.13083. Epub 2015 Jul 20. Epilepsia. 2015. PMID: 26190150
-
Seizure detection using scalp-EEG.Epilepsia. 2018 Jun;59 Suppl 1:14-22. doi: 10.1111/epi.14052. Epilepsia. 2018. PMID: 29873826 Review.
-
A new era in electroencephalographic monitoring? Subscalp devices for ultra-long-term recordings.Epilepsia. 2020 Sep;61(9):1805-1817. doi: 10.1111/epi.16630. Epub 2020 Aug 27. Epilepsia. 2020. PMID: 32852091 Review.
Cited by
-
Feasibility, Safety, and Performance of Full-Head Subscalp EEG Using Minimally Invasive Electrode Implantation.Neurology. 2024 Jun 25;102(12):e209428. doi: 10.1212/WNL.0000000000209428. Epub 2024 Jun 6. Neurology. 2024. PMID: 38843489 Free PMC article.
-
Imagined speech classification exploiting EEG power spectrum features.Med Biol Eng Comput. 2024 Aug;62(8):2529-2544. doi: 10.1007/s11517-024-03083-2. Epub 2024 Apr 18. Med Biol Eng Comput. 2024. PMID: 38632207
-
Novel subscalp and intracranial devices to wirelessly record and analyze continuous EEG in unsedated, behaving dogs in their natural environments: A new paradigm in canine epilepsy research.Front Vet Sci. 2022 Oct 20;9:1014269. doi: 10.3389/fvets.2022.1014269. eCollection 2022. Front Vet Sci. 2022. PMID: 36337210 Free PMC article. Review.
-
Optimizing Electrode Configurations for Wearable EEG Seizure Detection Using Machine Learning.Sensors (Basel). 2023 Jun 21;23(13):5805. doi: 10.3390/s23135805. Sensors (Basel). 2023. PMID: 37447653 Free PMC article.
-
The Individual Ictal Fingerprint: Combining Movement Measures With Ultra Long-Term Subcutaneous EEG in People With Epilepsy.Front Neurol. 2021 Dec 23;12:718329. doi: 10.3389/fneur.2021.718329. eCollection 2021. Front Neurol. 2021. PMID: 35002910 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical