Case Reports
. 2021 Apr 12;12(1):169-173.
doi: 10.1159/000510930. eCollection 2021 Jan-Apr.

Transient Clouding of a Sutureless Scleral Fixated Hydrophilic Intraocular Lens with Spontaneous Resolution: A Case Report and in vitro Experimental Test

Affiliations
Case Reports

Transient Clouding of a Sutureless Scleral Fixated Hydrophilic Intraocular Lens with Spontaneous Resolution: A Case Report and in vitro Experimental Test

Carla Danese et al. Case Rep Ophthalmol. .

Abstract

A 83-year-old woman underwent phacoemulsification in the left eye, complicated by posterior capsular rupture and dislocation of nuclear fragments in the vitreous cavity. She was left aphakic and referred to the Clinica Oculistica, Azienda Sanitaria Universitaria Friuli Centrale, Udine, Italy. After 2 days, she underwent pars plana vitrectomy, with removal of the nuclear fragments, and secondary intraocular lens (IOL) implantation. The lens used was a sutureless scleral fixation Carlevale lens (I71 FIL SSF Carlevale lens™. Soleko IOL Division), which was delivered to the operating theatre on the morning of the operation. During the implantation, it appeared opaque, without any deposit on its surfaces. The lens was not removed. On the next day, the lens was transparent, and it remained clear at the following control visits. An in vitro experimental test demonstrated that 25% hydrophilic polymers are vulnerable to thermal shock, developing an opacification which resolves after temperature stabilization of the lens.

Keywords: Cataract surgery; Hydrophilic intraocular lens; In vitro test; Intraocular lens; Transient intraocular lens clouding.

PubMed Disclaimer

Conflict of interest statement

Carla Danese: none. Renzo Calabresi: Soleko, E. Paolo Lanzetta: Bayer, C; Centervue, C; Novartis Pharma AG, C.

Figures

Fig. 1
Fig. 1
a Before implantation, the optic plate of the IOL appeared milky white. b The pupillary expansion ring allowed better visualization of the dense white opacity. c On the first post-operative day, the optic plate of the IOL appeared transparent. IOL, intraocular lens.
Fig. 2
Fig. 2
Results of the in vitro experimental test. a Opacification of the IOL at the moment of the termic shock. b–g Pictures showing the resolution of the clouding with the IOL being kept at constant temperature of 36°C (half an hour (b), one hour (c), one and a half hour (d), two hours (e), two and a half hours (f), three hours (g)). IOL, intraocular lens.

Similar articles

Cited by

References

    1. Neuhann T, Yildirim TM, Son HS, Merz PR, Khoramnia R, Auffarth GU. Reasons for explantation, demographics, and material analysis of 200 intraocular lens explants. J Cataract Refract Surg. 2020 Jan;46((1)):20–6. - PubMed
    1. McKibbin M, Seemongal-Dass RR, Atkinson PL. Transient fogging of acrylic (Acrysof) intraocular lenses. Eye (Lond) 1999 Oct;13((Pt 5)):672–3. - PubMed
    1. Tyagi P, Shah N, Jabir M. Intraoperative clouding of a posterior chamber intraocular lens. Int Ophthalmol. 2011 Dec;31((6)):483–4. - PubMed
    1. Helvacı S. Acute opacification of hydrophobic acrylic intraocular lens during implantation: result of temperature variation. Arq Bras Oftalmol. 2015 Jul-Aug;78((4)):267. - PubMed
    1. Zhang T, Li S, Liu C, Zhao R, Chang C, Han N. Acute clouding of a trifocal intraocular lens with spontaneous resolution: a case report. BMC Ophthalmol. 2019 Oct;19((1)):208. - PMC - PubMed

Publication types