Transient Clouding of a Sutureless Scleral Fixated Hydrophilic Intraocular Lens with Spontaneous Resolution: A Case Report and in vitro Experimental Test
- PMID: 33976677
- PMCID: PMC8077362
- DOI: 10.1159/000510930
Transient Clouding of a Sutureless Scleral Fixated Hydrophilic Intraocular Lens with Spontaneous Resolution: A Case Report and in vitro Experimental Test
Abstract
A 83-year-old woman underwent phacoemulsification in the left eye, complicated by posterior capsular rupture and dislocation of nuclear fragments in the vitreous cavity. She was left aphakic and referred to the Clinica Oculistica, Azienda Sanitaria Universitaria Friuli Centrale, Udine, Italy. After 2 days, she underwent pars plana vitrectomy, with removal of the nuclear fragments, and secondary intraocular lens (IOL) implantation. The lens used was a sutureless scleral fixation Carlevale lens (I71 FIL SSF Carlevale lens™. Soleko IOL Division), which was delivered to the operating theatre on the morning of the operation. During the implantation, it appeared opaque, without any deposit on its surfaces. The lens was not removed. On the next day, the lens was transparent, and it remained clear at the following control visits. An in vitro experimental test demonstrated that 25% hydrophilic polymers are vulnerable to thermal shock, developing an opacification which resolves after temperature stabilization of the lens.
Keywords: Cataract surgery; Hydrophilic intraocular lens; In vitro test; Intraocular lens; Transient intraocular lens clouding.
Copyright © 2021 by S. Karger AG, Basel.
Conflict of interest statement
Carla Danese: none. Renzo Calabresi: Soleko, E. Paolo Lanzetta: Bayer, C; Centervue, C; Novartis Pharma AG, C.
Figures
References
-
- Neuhann T, Yildirim TM, Son HS, Merz PR, Khoramnia R, Auffarth GU. Reasons for explantation, demographics, and material analysis of 200 intraocular lens explants. J Cataract Refract Surg. 2020 Jan;46((1)):20–6. - PubMed
-
- McKibbin M, Seemongal-Dass RR, Atkinson PL. Transient fogging of acrylic (Acrysof) intraocular lenses. Eye (Lond) 1999 Oct;13((Pt 5)):672–3. - PubMed
-
- Tyagi P, Shah N, Jabir M. Intraoperative clouding of a posterior chamber intraocular lens. Int Ophthalmol. 2011 Dec;31((6)):483–4. - PubMed
-
- Helvacı S. Acute opacification of hydrophobic acrylic intraocular lens during implantation: result of temperature variation. Arq Bras Oftalmol. 2015 Jul-Aug;78((4)):267. - PubMed