Ultrasound Monitoring of Umbilical Catheters in the Neonatal Intensive Care Unit-A Prospective Observational Study
- PMID: 33996700
- PMCID: PMC8119780
- DOI: 10.3389/fped.2021.665214
Ultrasound Monitoring of Umbilical Catheters in the Neonatal Intensive Care Unit-A Prospective Observational Study
Abstract
Introduction: Umbilical catheterization provides a quick yet demanding central line that can lead to complications seen nowhere else. The aim of our study was to determine whether the repeated ultrasound scanning can influence the catheterization time, prevent some of the catheter-related complications, support the decision-making process and allow prolonged catheterization in patients without an alternative central access route. Methods: A prospective observational study was performed in a tertiary neonatal intensive care unit. A total of 129 patients and 194 umbilical catheters (119 venous and 75 arterial) were analyzed with a total of 954 scans. Ultrasound screening consisted of 1) assessing the catheter tip, location, movability, and surface and 2) analyzing the catheter trajectory. The outcome variables were defined as 1) catheter dislocation and 2) associated thrombosis. Results: Dislocation of catheter throughout the whole catheterization period was observed in 68% (81/119) of UVCs and 23% (17/75) of UACs. Thrombotic complications were observed in 34.5% (41/119) of UVCs and 12% (9/75) of UACs. 1/3 of UAC-associated thrombi were visible only after catheter removal. 51% (61/119) of UVC patients and 8% (6/75) of UAC patients made a clinical decision regarding the obtained catheter image. Conclusion: Bedside ultrasound imaging of catheters supports the decision-making process related to the catheterization duration, shortening the time if abnormalities are detected and allowing a safer prolonged UC stay when an alternative central line cannot be inserted.
Keywords: bedside monitoring; catheter-related thrombosis; neonatal intensive care unit; prematurity; umbilical catheters.
Copyright © 2021 Sobczak, Dudzik, Kruczek and Kwinta.
Conflict of interest statement
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Figures
References
-
- Singh Y, Tissot C, Fraga MV, Yousef N, Cortes RG, Lopez J, et al. . International evidence-based guidelines on Point of Care Ultrasound (POCUS) for critically ill neonates and children issued by the POCUS Working Group of the European Society of Paediatric and Neonatal Intensive Care (ESPNIC). Crit Care. (2020) 24:1–16. 10.1186/s13054-020-2787-9 - DOI - PMC - PubMed
-
- WHO . Preterm Birth. Fact Sheet; (2014).
-
- Oski F, Allen D, Diamond L. Portal hypertension a complication of umbilical vein catheterization. Pediatrics. (1963) 31:297–302. - PubMed