Comparison of Transmittance and Reflectance Pulse Oximetry in Anesthetized Dogs
- PMID: 33996971
- PMCID: PMC8119740
- DOI: 10.3389/fvets.2021.643966
Comparison of Transmittance and Reflectance Pulse Oximetry in Anesthetized Dogs
Abstract
Objectives: The tongue is the standard site for placement of a pulse oximeter probe but is difficult to access during certain procedures such as dental and ophthalmic procedures and computerized tomography of the head. The aim of this study was to evaluate the performance of a new-generation reflectance pulse oximeter using the tail and tibia as sites for probe attachment. Materials and Methods: A total of 100 client-owned dogs that underwent anesthesia for various reasons were premedicated with butorphanol (n = 50; 0.2 mg/kg; group BUT) or butorphanol and dexmedetomidine (n = 50; 5 μg/kg; group DEX), administered intravenously. Anesthesia was induced with propofol and maintained with sevoflurane. A transmittance pulse oximeter probe was placed on the tongue and served as the reference standard. A reflectance probe was randomly placed on the tail base or the proximal tibia, and the position changed after testing. Signals from three consecutive measurements were obtained at each position. Failure was defined as "no signal," "low signal," or a pulse difference >10/min compared with the ECG heart rate. Data were analyzed using chi-square test, Wilcoxon matched-pair signed-rank test, and Bland-Altman analysis. P < 0.05 was considered significant. Results: In both groups (BUT and DEX), failure rate was higher when the tibia and tail were used as probe sites compared with the tongue. In both groups, the failure rate was higher for the tibia than for the tail. Dexmedetomidine-induced vasoconstriction increased failure rate at all probe positions. Clinical Significance: The tail base, but not the tibia, is an acceptable position for reflectance pulse oximeter probes in dogs. The tongue remains the probe site of choice, if accessible.
Keywords: canine; hypoxia; oxygen saturation; tail; tibia tongue.
Copyright © 2021 Nixdorff, Zablotski, Hartmann and Dörfelt.
Conflict of interest statement
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Similar articles
-
Comparison of human smartwatch and transmittance pulse oximetry for evaluating peripheral oxygen saturation in anesthetized dogs.Iran J Vet Res. 2024;25(2):166-169. doi: 10.22099/IJVR.2024.47628.6884. Iran J Vet Res. 2024. PMID: 39624191 Free PMC article.
-
Evaluation of the performance of two new generation pulse oximeters in cats at different probe positions and under the influence of vasoconstriction.J Feline Med Surg. 2022 Oct;24(10):1026-1031. doi: 10.1177/1098612X211063768. Epub 2021 Dec 14. J Feline Med Surg. 2022. PMID: 34904479 Free PMC article.
-
Evaluation of pulse oximeter probes and probe placement in healthy dogs.J Am Anim Hosp Assoc. 1995 Jan-Feb;31(1):9-14. doi: 10.5326/15473317-31-1-9. J Am Anim Hosp Assoc. 1995. PMID: 7820772
-
Experimental and clinical evaluation of a noninvasive reflectance pulse oximeter sensor.J Clin Monit. 1992 Oct;8(4):257-66. doi: 10.1007/BF01617907. J Clin Monit. 1992. PMID: 1453185
-
Evaluation of accuracy of pulse oximetry in dogs.Am J Vet Res. 1992 Apr;53(4):537-40. Am J Vet Res. 1992. PMID: 1586024
Cited by
-
The utility of the respiratory rate-oxygenation index as a predictor of treatment response in dogs receiving high-flow nasal cannula oxygen therapy.Front Vet Sci. 2024 May 7;11:1404195. doi: 10.3389/fvets.2024.1404195. eCollection 2024. Front Vet Sci. 2024. PMID: 38774907 Free PMC article.
-
Use of Ketamine or Xylazine to Provide Balanced Anesthesia with Isoflurane in C57BL/6J Mice.J Am Assoc Lab Anim Sci. 2022 Sep 1;61(5):457-467. doi: 10.30802/AALAS-JAALAS-21-000125. Epub 2022 Aug 8. J Am Assoc Lab Anim Sci. 2022. PMID: 35940848 Free PMC article.
-
Evaluation of the performance of three pulse oximeters at different probe positions in awake rabbits.PLoS One. 2025 May 15;20(5):e0323044. doi: 10.1371/journal.pone.0323044. eCollection 2025. PLoS One. 2025. PMID: 40373047 Free PMC article.
-
Noninvasive hemoglobin sensing and imaging: optical tools for disease diagnosis.J Biomed Opt. 2022 Aug;27(8):080901. doi: 10.1117/1.JBO.27.8.080901. J Biomed Opt. 2022. PMID: 35922891 Free PMC article.
-
Comparison of human smartwatch and transmittance pulse oximetry for evaluating peripheral oxygen saturation in anesthetized dogs.Iran J Vet Res. 2024;25(2):166-169. doi: 10.22099/IJVR.2024.47628.6884. Iran J Vet Res. 2024. PMID: 39624191 Free PMC article.
References
-
- Mendelson Y, Kent JC, Yocum BL, Birle MJ. Design and evaluation of a new reflectance pulse oximeter sensor. Med Instrum. (1988) 22:167–73. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources