Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Practice Guideline
. 2022 Apr 1;146(4):440-450.
doi: 10.5858/arpa.2020-0723-CP.

Validating Whole Slide Imaging Systems for Diagnostic Purposes in Pathology

Affiliations
Free article
Practice Guideline

Validating Whole Slide Imaging Systems for Diagnostic Purposes in Pathology

Andrew J Evans et al. Arch Pathol Lab Med. .
Free article

Abstract

Context.—: The original guideline, "Validating Whole Slide Imaging for Diagnostic Purposes in Pathology," was published in 2013 and included 12 guideline statements. The College of American Pathologists convened an expert panel to update the guideline following standards established by the National Academies of Medicine for developing trustworthy clinical practice guidelines.

Objective.—: To assess evidence published since the release of the original guideline and provide updated recommendations for validating whole slide imaging (WSI) systems used for diagnostic purposes.

Design.—: An expert panel performed a systematic review of the literature. Frozen sections, anatomic pathology specimens (biopsies, curettings, and resections), and hematopathology cases were included. Cytology cases were excluded. Using the Grading of Recommendations Assessment, Development, and Evaluation approach, the panel reassessed and updated the original guideline recommendations.

Results.—: Three strong recommendations and 9 good practice statements are offered to assist laboratories with validating WSI digital pathology systems.

Conclusions.—: Systematic review of literature following release of the 2013 guideline reaffirms the use of a validation set of at least 60 cases, establishing intraobserver diagnostic concordance between WSI and glass slides and the use of a 2-week washout period between modalities. Although all discordances between WSI and glass slide diagnoses discovered during validation need to be reconciled, laboratories should be particularly concerned if their overall WSI-glass slide concordance is less than 95%.