. 2022 Jan 1;48(1):26-31.
doi: 10.1097/j.jcrs.0000000000000704.

Comparative analysis of 2 biometers using swept-source OCT technology

Affiliations

Comparative analysis of 2 biometers using swept-source OCT technology

Christophe Panthier et al. J Cataract Refract Surg. .

Abstract

Purpose: To report the repeatability, level of agreement, and correlation of 2 swept-source optical coherence tomography (SS-OCT) biometers, the IOLMaster 700 (biometer A) and the ANTERION (biometer B).

Setting: Rothschild Foundation Hospital, Paris, France.

Design: Prospective comparative case series.

Methods: Consecutive patients who attended for cataract assessment had SS-OCT biometry performed with the reference, biometer A, and biometer B. Axial length, mean keratometry, central corneal thickness, anterior chamber depth, lens thickness, and corneal diameter measurements (white-to-white [WTW] distance) were measured. The intraoperator repeatability was calculated using analysis of variance and repeatability limits. Correlations were assessed, and the level of agreement between the biometers was represented with the Bland-Altman method.

Results: The study comprised 63 patients (125 eyes). There was a statistically significant difference between the 2 biometers in all measurements (P < .05). All repeatability measurements were very high, as were the level of correlation and level of agreement between biometer A and biometer B for all parameters.

Conclusions: Biometer B provided good agreement and repeatability compared with biometer A; however, all parameters were not interchangeable, particularly, the WTW measurement and the high keratometric values.

PubMed Disclaimer

References

    1. Hirnschall N, Varsits R, Doeller B, Findl O. Enhanced penetration for axial length measurement of eyes with dense cataracts using swept source optical coherence tomography: a consecutive observational study. Ophthalmol Ther2018;7:119–124
    1. Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography–based optical biometer and a time-domain optical coherence tomography–based optical biometer. J Cataract Refract Surg2015;41:2224–2232
    1. Hirnschall N, Murphy S, Pimenides D, Maurino V, Findl O. Assessment of a new averaging algorithm to increase the sensitivity of axial eye length measurement with optical biometry in eyes with dense cataract. J Cataract Refract Surg2011;37:45–49
    1. Akman A, Asena L, Güngör SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol2016;100:1201–1205
    1. Findl O, Kriechbaum K, Sacu S, Kiss B, Polak K, Nepp J, Schild G, Rainer G, Maca S, Petternel V, Lackner B, Drexler W. Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery. J Cataract Refract Surg 2003;29:1950–1955

LinkOut - more resources