Alarms in a neurocritical care unit: a prospective study

Affiliations

Alarms in a neurocritical care unit: a prospective study

Ali Unal et al. J Clin Monit Comput. 2022 Aug.

Abstract

The contemporary practice of monitoring physiologic parameters in the critical care setting is based on alarm systems with high sensitivity but low specificity. A natural consequence of this approach is a massive amount of alarms, which potentially leads to fatigue in the personnel and negatively impacts the quality of care provided. The study objective is to determine the prevalence, types, and determinants of alarms in a neurological critical care unit (NCCU) prototype. During a one-month period corresponding to 272 days of monitoring in 34 patients, nursing staff recorded the type and number of sounding alarms in a university NCCU. Alarms were categorized into three types as type-A alarms that were merely handled by the nursing staff, type-B alarms that were primarily managed by nurses, but the physician was also notified, and type-C alarms that were principally handled by NCCU physicians. There were a total of 9439 alarms, with an average of daily 34.7 alarms per bed, corresponding to one alarm every 41.4 min. Most of the alarms were type-A (57.7%), followed by type-B (39.2%) and type-C (3.1%) alarms. Alarms originated from electrocardiogram (34.6%), pulse oximeter (33.7%), noninvasive blood pressure monitoring (9.8%), respiratory monitoring (9.7%), intravenous fluid pumps (4.5%), ventilator (3.9%), enteral pumps (2.1%) and invasive blood pressure systems (1.7%). A noticeable diurnal variation was observed for type-A pulse oximeter, type-A and -B ECG alarms (increase during morning shifts), and type-A ventilator alarms (decrease during morning shifts). Alarms are highly prevalent in NCCUs and can correspond to an important portion of the workload.

Keywords: Alarm; Alarm fatigue; Artificial intelligence; False-negative; Neurointensive care; Prognosis; Specialty.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no conflict of interest.

Figures

Fig. 1
Fig. 1
Diurnal variation of alarm subtypes. ECG electrocardiogram, IBP invasive blood pressure, IV intravenous, MV mechanical ventilation, NIBP non-invasive blood pressure, RR respiratory rate, SO2 oxygen saturation. *p < 0.05 (significant), +p < 0.1 (trend)

References

    1. Phillips J. Clinical alarms: complexity and common sense. Crit Care Nurs Clin North Am. 2006;18(2):145–156. doi: 10.1016/j.ccell.2006.01.002. - DOI - PubMed
    1. Lawson N, Thompson K, Saunders G, Saiz J, Richardson J, Brown D, Ince N, Caldwell M, Pope D.Sound intensity and noise evaluation in a critical care unit. Am J Crit Care. 2010;19(6):e88–98; quiz e99. 10.4037/ajcc2010180. - PubMed
    1. Graham KC, Cvach M. Monitor alarm fatigue: standardizing use of physiological monitors and decreasing nuisance alarms. Am J Crit Care. 2010;19(1):28–34;quiz 35. 10.4037/ajcc2010651. - PubMed
    1. Wilken M, Huske-Kraus D, Rohrig R. Alarm fatigue: using alarm data from a patient data monitoring system on an intensive care unit to improve the alarm management. Stud Health Technol Inform. 2019;267:273–281. doi: 10.3233/SHTI190838. - DOI - PubMed
    1. Wung SF, Schatz MR. Critical care nurses' cognitive ergonomics related to medical device alarms. Crit Care Nurs Clin North Am. 2018;30(2):191–202. doi: 10.1016/j.cnc.2018.02.002. - DOI - PubMed

LinkOut - more resources