Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 May 19;21(10):3545.
doi: 10.3390/s21103545.

A Crosstalk- and Interferent-Free Dual Electrode Amperometric Biosensor for the Simultaneous Determination of Choline and Phosphocholine

Affiliations

A Crosstalk- and Interferent-Free Dual Electrode Amperometric Biosensor for the Simultaneous Determination of Choline and Phosphocholine

Rosanna Ciriello et al. Sensors (Basel). .

Abstract

Choline (Ch) and phosphocholine (PCh) levels in tissues are associated to tissue growth and so to carcinogenesis. Till now, only highly sophisticated and expensive techniques like those based on NMR spectroscopy or GC/LC- high resolution mass spectrometry permitted Ch and PCh analysis but very few of them were capable of a simultaneous determination of these analytes. Thus, a never reported before amperometric biosensor for PCh analysis based on choline oxidase and alkaline phosphatase co-immobilized onto a Pt electrode by co-crosslinking has been developed. Coupling the developed biosensor with a parallel sensor but specific to Ch, a crosstalk-free dual electrode biosensor was also developed, permitting the simultaneous determination of Ch and PCh in flow injection analysis. This novel sensing device performed remarkably in terms of sensitivity, linear range, and limit of detection so to exceed in most cases the more complex analytical instrumentations. Further, electrode modification by overoxidized polypyrrole permitted the development of a fouling- and interferent-free dual electrode biosensor which appeared promising for the simultaneous determination of Ch and PCh in a real sample.

Keywords: alkaline phosphatase; choline analysis; choline oxidase; dual electrode biosensor; electropolymerized non-conducting polymer; enzyme immobilization; flow injection analysis; overoxidized polypyrrole; phosphocholine analysis; simultaneous determination.

PubMed Disclaimer

Conflict of interest statement

The authors declare no conflict of interest.

Figures

Scheme 1
Scheme 1
The approach for the simultaneous determination by the dual electrode (parallel configuration) amperometric biosensor thin-layer cell (as shown in Scheme S1 in Supplementary Materials) used in the flow injection experiments for Ch and PCh analysis (a) and the example of a crosstalk effect in measurements (b). ChO and ALP/ChO refer to ChO and ALP/ChO biosensors, respectively, the black curved arrows (left side in both sketches) represent the enzymatic detection, the blue ones, the carrier flow direction while iCh and iPCh the currents for Ch and PCh measurements, respectively, as measured at their respective electrical connections. The larger black curved arrow (right side in right sketch) represents the potential crosstalk effect due to hydrogen peroxide detection from the side electrode.
Figure 1
Figure 1
Typical flow injection responses at the dual electrode biosensor for replicate injections of Ch 0.1 mM (Ch1), 0.05 mM (Ch2), 0.025 mM (Ch3), 0.01 mM (Ch4) and PCh 0.5 mM (PCh1), 0.25 mM (PCh2), 0.1 mM (PCh3), 0.05 mM (PCh4). Upper and lower traces refer to responses at Pt/ChO and Pt/Cho-ALP biosensors, respectively. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in Section 2.
Figure 2
Figure 2
Calibration curves at the dual electrode biosensor for replicate injections of Ch and PCh. Continuous lines refer to Michaelis-Menten fitting of data. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in the Section 2.
Figure 3
Figure 3
Typical flow injection responses at the dual electrode biosensor for replicate injections of Ch 2.5 µM (Ch1), 1.25 µM (Ch2), 0.625 µM (Ch3) and PCh 25 µM (PCh1), 12.5 µM (PCh2), 6.25 µM (PCh3); B refers to responses due to replicate injections of the carrier solution, i.e., the buffer used for preparing sample solutions. Upper and lower traces refer to responses at Pt/ChO and Pt/ChO-ALP biosensors, respectively. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in Section 2.
Figure 4
Figure 4
Calibration 3d plot at the Pt/ChO electrode of the dual electrode biosensor for replicate injections of mixtures of Ch and PCh. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in Section 2.
Figure 5
Figure 5
Calibration lines at the Pt/ChO electrode of the dual electrode biosensor for replicate injections of mixtures of Ch and PCh at several concentration levels of PCh (upper plot) and Ch (lower plot). Continuous lines refer to linear fitting of data. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described Section 2.
Figure 6
Figure 6
Calibration 3d plot at the Pt/ChO-ALP electrode of the dual electrode biosensor for replicate injections of mixtures of Ch and PCh. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in Section 2.
Figure 7
Figure 7
Calibration lines at the Pt/ChO-ALP electrode of the dual electrode biosensor for replicate injections of mixtures of Ch and PCh at several concentration levels of PCh (upper plot) and Ch (lower plot). Continuous lines refer to linear fitting of data. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in Section 2.
Figure 8
Figure 8
Typical flow injection responses at the oPPy modified dual electrode biosensor for replicate injections of Ch 2.5 µM (Ch1), 1.25 µM (Ch2), 0.625 µM (Ch3), PCh 25 µM (PCh1), 12.5 µM (PCh2), 6.25 µM (PCh3) compared to responses due to AA 0.1 mM, AU 0.5 mM, and CYS 0.2 mM; B refers to responses due to replicate injections of the carrier solution, i.e., the buffer used for preparing sample solutions. Upper and lower traces refer to responses at Pt/oPPy/ChO and Pt/oPPy/ChO-ALP biosensors, respectively. Carrier solution: borate buffer (pH 9, I 0.1 M); flow rate 1 mL/min; injection volume 20 µL. Other experimental conditions as described in Section 2.

Similar articles

Cited by

References

    1. Zeisel S.H. Choline deficiency. J. Nutr. Biochem. 1990;1:332–349. doi: 10.1016/0955-2863(90)90001-2. - DOI - PubMed
    1. Kiss Z. Regulation of Mitogenesis by Water-Soluble Phospholipid Intermediates1Abbreviations: PCho—phosphocholine; PEtn—phosphoethanolamine; PtdChoߞphosphatidylcholine; PtdEtn—phosphatidylethanolamine; CK—choline kinase; EK—ethanolamine kinase; PLD—phospholipase D. Cell. Signal. 1999;11:149–157. doi: 10.1016/S0898-6568(98)00057-6. - DOI - PubMed
    1. Katz-Brull R., Margalit R., Bendel P., Degani H. Choline metabolism in breast cancer; H-, C- and P-NMR studies of cells and tumors. Magn. Reson. Mater. Biol. Phys. Med. 1998;6:44–52. doi: 10.1007/BF02662511. - DOI - PubMed
    1. Negendank W. Studies of human tumors by MRS: A review. NMR Biomed. 1992;5:303–324. doi: 10.1002/nbm.1940050518. - DOI - PubMed
    1. Sharma U., Mehta A., Seenu V., Jagannathan N.R. Biochemical characterization of metastatic lymph nodes of breast cancer patients by in vitro 1H magnetic resonance spectroscopy: A pilot study. Magn. Reson. Imaging. 2004;22:697–706. doi: 10.1016/j.mri.2004.01.037. - DOI - PubMed

LinkOut - more resources