Bacterial colonization and phlebitis-associated risk with transparent polyurethane film for peripheral intravenous site dressings
- PMID: 3408013
- DOI: 10.1016/0196-6553(88)90046-6
Bacterial colonization and phlebitis-associated risk with transparent polyurethane film for peripheral intravenous site dressings
Abstract
Previous studies of various brands of polyurethane dressings have noted differences in the rates of catheter colonization. We compared Bioclusive transparent polyurethane (TP) dressing with a cotton gauze (CG) dressing on peripheral intravenous (IV) access sites for the incidence of phlebitis, catheter tip colonization, skin colonization, and catheter-related bacteremia. The study, involving 598 ward patients, was case controlled, prospective, and randomized for a period of 4 months. Each patient was entered into the study only once, and all dressings were applied by a member of the IV therapy team. No significant difference was seen for phlebitis rate (TP: 9.8% vs. CG: 7.6%) or catheter tip colonization, defined as greater than 15 colony forming units (CFU) (5.7% vs. 4.4%) by a semiquantitative technique. Cultures of specimens from the skin and catheter tips of the majority of patients (91%) showed no growth. An association was found between those patients with greater than 15 CFU isolated from catheter tips and those with phlebitis (p = 0.022). No documented catheter-related bacteremia occurred in either study group.