Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Jul;22(7):147-154.
doi: 10.1002/acm2.13293. Epub 2021 Jun 7.

Impact of errors in spot size and spot position in robustly optimized pencil beam scanning proton-based stereotactic body radiation therapy (SBRT) lung plans

Affiliations

Impact of errors in spot size and spot position in robustly optimized pencil beam scanning proton-based stereotactic body radiation therapy (SBRT) lung plans

Suresh Rana et al. J Appl Clin Med Phys. 2021 Jul.

Abstract

Purpose: The purpose of the current study was threefold: (a) investigate the impact of the variations (errors) in spot sizes in robustly optimized pencil beam scanning (PBS) proton-based stereotactic body radiation therapy (SBRT) lung plans, (b) evaluate the impact of spot sizes and position errors simultaneously, and (c) assess the overall effect of spot size and position errors occurring simultaneously in conjunction with either setup or range errors.

Methods: In this retrospective study, computed tomography (CT) data set of five lung patients was selected. Treatment plans were regenerated for a total dose of 5000 cGy(RBE) in 5 fractions using a single-field optimization (SFO) technique. Monte Carlo was used for the plan optimization and final dose calculations. Nominal plans were normalized such that 99% of the clinical target volume (CTV) received the prescription dose. The analysis was divided into three groups. Group 1: The increasing and decreasing spot sizes were evaluated for ±10%, ±15%, and ±20% errors. Group 2: Errors in spot size and spot positions were evaluated simultaneously (spot size: ±10%; spot position: ±1 and ±2 mm). Group 3: Simulated plans from Group 2 were evaluated for the setup (±5 mm) and range (±3.5%) errors.

Results: Group 1: For the spot size errors of ±10%, the average reduction in D99% for -10% and +10% errors was 0.7% and 1.1%, respectively. For -15% and +15% spot size errors, the average reduction in D99% was 1.4% and 1.9%, respectively. The average reduction in D99% was 2.1% for -20% error and 2.8% for +20% error. The hot spot evaluation showed that, for the same magnitude of error, the decreasing spot sizes resulted in a positive difference (hotter plan) when compared with the increasing spot sizes. Group 2: For a 10% increase in spot size in conjunction with a -1 mm (+1 mm) shift in spot position, the average reduction in D99% was 1.5% (1.8%). For a 10% decrease in spot size in conjunction with a -1 mm (+1 mm) shift in spot position, the reduction in D99% was 0.8% (0.9%). For the spot size errors of ±10% and spot position errors of ±2 mm, the average reduction in D99% was 2.4%. Group 3: Based on the results from 160 plans (4 plans for spot size [±10%] and position [±1 mm] errors × 8 scenarios × 5 patients), the average D99% was 4748 cGy(RBE) with the average reduction of 5.0%. The isocentric shift in the superior-inferior direction yielded the least homogenous dose distributions inside the target volume.

Conclusion: The increasing spot sizes resulted in decreased target coverage and dose homogeneity. Similarly, the decreasing spot sizes led to a loss of target coverage, overdosage, and degradation of dose homogeneity. The addition of spot size and position errors to plan robustness parameters (setup and range uncertainties) increased the target coverage loss and decreased the dose homogeneity.

Keywords: Monte Carlo; SBRT; lung cancer; proton therapy; robust optimization; robustness; spot position; spot size.

PubMed Disclaimer

Conflict of interest statement

The authors do not have any relevant conflict of interest to disclose.

Figures

Fig. 1
Fig. 1
The average difference in clinical target volume D95% (left panel) and D99% (right panel) between simulated plans in Group 1 and nominal plans for the spot size errors (±10%, ±15%, and ±20%). The results are averaged over all five patients.
Fig. 2
Fig. 2
Dose distributions in an example patient: (a) nominal plan, (b) simulated plan for the decreasing spot size (−10%) and spot position (+1 mm) evaluated simultaneously; (c) simulated plan for the increasing spot size (+10%) and spot position (+1 mm) evaluated simultaneously. The loss of target coverage in the simulated plans is shown by the red arrows on the right panel.
Fig. 3
Fig. 3
The average difference in clinical target volume D99% between simulated plans in Group 2 and nominal plans for the spot size errors (±10%) in conjunction with spot position errors (±1 and ±2 mm). The results are averaged over all five patients.
Fig. 4
Fig. 4
(left panel) The average clinical target volume (CTV) D99% (left panel) from 120 plans of five patients from Group 3 analysis. (right panel) The average difference in CTV D99% between simulated plans in Group 3 and nominal plans. The results are averaged over all five patients.
Fig. 5
Fig. 5
(left panel) The difference in clinical target volume homogeneity index (HI) for eight different scenarios between simulated plans in Group 3 and nominal plans. (right panel) Robustness parameters for eight scenarios of Group 3 analysis. The D(±10%, ±1 mm) plans are evaluated for setup (±5 mm) and range (±3.5%) uncertainties as described in Section 2.F.

References

    1. Parodi K, Mairani A, Brons S, et al. The influence of lateral beam profile modifications in scanned proton and carbon ion therapy: a Monte Carlo study. Phys Med Biol. 2010;55:5169–5187. - PubMed
    1. Rana S, Rosenfeld AB. Parametrization of in‐air spot size as a function of energy and air gap for the ProteusPLUS pencil beam scanning proton therapy system. Radiol Phys Technol. 2020;13:392–397. - PubMed
    1. Peterson S, Polf J, Ciangaru G, Frank SJ, Bues M, Smith A. Variations in proton scanned beam dose delivery due to uncertainties in magnetic beam steering. Med Phys. 2009;36:3693–3702. - PubMed
    1. Yu J, Beltran CJ, Herman MG. Implication of spot position error on plan quality and patient safety in pencil‐beam‐scanning proton therapy. Med Phys. 2014;41:081706. - PubMed
    1. Chanrion MA, Ammazzalorso F, Wittig A, Engenhart‐Cabillic R, Jelen U. Dosimetric consequences of pencil beam width variations in scanned beam particle therapy. Phys Med Biol. 2013;58:3979–3993. - PubMed