Standard versus strict stability criteria in radiofrequency paroxysmal atrial fibrillation ablation using ablation index
- PMID: 34224159
- DOI: 10.1111/pace.14309
Standard versus strict stability criteria in radiofrequency paroxysmal atrial fibrillation ablation using ablation index
Abstract
Purpose: To compare the outcome of paroxysmal atrial fibrillation (AF) ablation via pulmonary vein isolation using Ablation Index (AI) with strict or standard stability criteria.
Methods: We enrolled 130 consecutive naive patients affected by paroxysmal AF who underwent PVI at two high-volume centers. AI target was ≥380 at the posterior wall and ≥500 at the anterior wall. Strict versus standard stability criteria were set for Group 1 (65 patients) and Group 2 (65 patients), respectively. We compared those strategies with a historical cohort of 72 consecutive patients treated at same centers in the VISITALY study, using average force ≥10 g and strict stability criteria as target parameters. Interlesion distance target was <6 mm. Recurrence was defined as any AF, atrial tachycardia (AT) or atrial flutter (AFL) during the 12 months after ablation, excluding a 90-days blanking period.
Results: Procedure duration (224.05 ± 47.21 vs. 175.61 ± 51.29 min; p < .001), fluoroscopy time (11.85 ± 4.38 vs. 10.46 ± 6.49 min; p = .019) and pericardial effusion rate (9.23% vs. 0%; p = .01) were higher in Group 1 than in Group 2. Freedom from AF/AT/AFL at 12 months was not significantly different (Group 1: 86.15%; Group 2: 90.77%; p = .42). Compared to VISITALY study, there were not significant differences in terms of recurrences.
Conclusion: A strategy of PVI using AI with standard stability criteria performed the best in terms of procedure efficiency and safety. Twelve-months arrhythmia-free survival rate was comparable with other strategies pursuing an interlesion distance target <6 mm, regardless of the use of AI.
© 2021 Wiley Periodicals LLC.
References
REFERENCES
-
- Kuck K-H, Brugada J, Fürnkranz A, et al. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med. 2016;374:2235-2245.
-
- Hindricks G, Potpara T, Dagres N, et al. 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association of Cardio-Thoracic Surgery (EACTS): the task force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2020;42:373-498.
-
- Park C-I, Lehrmann H, Keyl C, et al. Mechanisms of pulmonary vein reconnection after radiofrequency ablation of atrial fibrillation: the deterministic role of contact force and interlesion distance. J Cardiovasc Electrophysiol. 2014;25:701-708.
-
- Phlips T, Taghji P, El Haddad M, et al. Improving procedural and one-year outcome after contact force-guided pulmonary vein isolation: the role of interlesion distance, ablation index, and contact force variability in the ’CLOSE’-protocol. EP Europace. 2018;20:f419-f427.
-
- Zucchelli G, Sirico G, Rebellato L, et al. Contiguity between ablation lesions and strict catheter stability settings assessed by VISITAG™ module improve clinical outcomes of paroxysmal atrial fibrillation ablation: Results from the VISITALY study. Circ J. 2018;82:974-982.