Performance Characteristics of the Biograph Vision Quadra PET/CT System with a Long Axial Field of View Using the NEMA NU 2-2018 Standard
- PMID: 34301780
- DOI: 10.2967/jnumed.121.261972
Performance Characteristics of the Biograph Vision Quadra PET/CT System with a Long Axial Field of View Using the NEMA NU 2-2018 Standard
Abstract
Our purpose was to evaluate the performance of the Biograph Vision Quadra PET/CT system. This new system is based on the Biograph Vision 600, using the same silicon photomultiplier-based detectors with 3.2 × 3.2 × 20 mm lutetium-oxoorthosilicate crystals. The 32 detector rings of the Quadra provide a 4-fold larger axial field of view (AFOV) of 106 cm, enabling imaging of major organs in 1 bed position. Methods: The physical performance of the scanner was evaluated according to the National Electrical Manufacturers Association NU 2-2018 standard, with additional experiments to characterize energy resolution. Image quality was assessed with foreground-to-background ratios of 4:1 and 8:1. Additionally, a clinical 18F-FDG PET study was reconstructed with varying frame durations. In all experiments, data were acquired using the maximum ring distance of 322 crystals (MRD 322), whereas image reconstructions could be performed with a maximum ring distance of only 85 crystals (MRD 85). Results: The spatial resolution at full width at half maximum in the radial, tangential, and axial directions was 3.3, 3.4, and 3.8 mm, respectively. The sensitivity was 83 cps/kBq for MRD 85 and 176 cps/kBq for MRD 322. The noise-equivalent count rates (NECRs) at peak were 1,613 kcps for MRD 85 and 2,956 kcps for MRD 322, both at 27.49 kBq/mL. The respective scatter fractions at peak NECR equaled 36% and 37%. The time-of-flight resolution at peak NECR was 228 ps for MRD 85 and 230 ps for MRD 322. Image contrast recovery ranged from 69.6% to 86.9% for 4:1 contrast ratios and from 77.7% to 92.6% for 8:1 contrast ratios reconstructed using point-spread function time of flight with 8 iterations and 5 subsets. Thirty-second frames provided readable lesion detectability and acceptable noise levels in clinical images. Conclusion: The Biograph Vision Quadra PET/CT device has spatial and time resolution similar to those of the Biograph Vision 600 but exhibits improved sensitivity and NECR because of its extended AFOV. The reported spatial resolution, time resolution, and sensitivity make it a competitive new device in the class of PET scanners with an extended AFOV.
Keywords: NEMA; acceptance test; digital PET; long field of view; total-body.
© 2022 by the Society of Nuclear Medicine and Molecular Imaging.
Similar articles
-
Performance Characteristics of the Digital Biograph Vision PET/CT System.J Nucl Med. 2019 Jul;60(7):1031-1036. doi: 10.2967/jnumed.118.215418. Epub 2019 Jan 10. J Nucl Med. 2019. PMID: 30630944
-
Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards.J Nucl Med. 2024 Dec 3;65(12):1974-1982. doi: 10.2967/jnumed.124.267963. J Nucl Med. 2024. PMID: 39510588 Free PMC article.
-
Monte Carlo evaluation of hypothetical long axial field-of-view PET scanner using GE Discovery MI PET front-end architecture.Med Phys. 2022 Feb;49(2):1139-1152. doi: 10.1002/mp.15422. Epub 2022 Jan 10. Med Phys. 2022. PMID: 34954831
-
A review of digital PET-CT technology: Comparing performance parameters in SiPM integrated digital PET-CT systems.Radiography (Lond). 2024 Jan;30(1):13-20. doi: 10.1016/j.radi.2023.10.004. Epub 2023 Oct 19. Radiography (Lond). 2024. PMID: 37864986 Review.
-
Long-axial field-of-view PET/CT: perspectives and review of a revolutionary development in nuclear medicine based on clinical experience in over 7000 patients.Cancer Imaging. 2023 Mar 18;23(1):28. doi: 10.1186/s40644-023-00540-3. Cancer Imaging. 2023. PMID: 36934273 Free PMC article. Review.
Cited by
-
Application of the long axial field-of-view PET/CT with low-dose [18F]FDG in melanoma.Eur J Nucl Med Mol Imaging. 2023 Mar;50(4):1158-1167. doi: 10.1007/s00259-022-06070-7. Epub 2022 Dec 7. Eur J Nucl Med Mol Imaging. 2023. PMID: 36474125 Free PMC article.
-
16-channel SiPM high-frequency readout with time-over-threshold discrimination for ultrafast time-of-flight applications.EJNMMI Phys. 2023 Dec 4;10(1):76. doi: 10.1186/s40658-023-00594-z. EJNMMI Phys. 2023. PMID: 38044383 Free PMC article.
-
A review of harmonization strategies for quantitative PET.Ann Nucl Med. 2023 Feb;37(2):71-88. doi: 10.1007/s12149-022-01820-x. Epub 2023 Jan 6. Ann Nucl Med. 2023. PMID: 36607466 Free PMC article. Review.
-
Long Axial Field-of-View PET for Ultra-Low-Dose Imaging of Non-Hodgkin Lymphoma during Pregnancy.Diagnostics (Basel). 2022 Dec 22;13(1):28. doi: 10.3390/diagnostics13010028. Diagnostics (Basel). 2022. PMID: 36611320 Free PMC article.
-
Unsupervised deep learning framework for data-driven gating in positron emission tomography.Med Phys. 2023 Oct;50(10):6047-6059. doi: 10.1002/mp.16642. Epub 2023 Aug 4. Med Phys. 2023. PMID: 37538038 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials