. 2022 Feb 1;48(2):168-172.
doi: 10.1097/j.jcrs.0000000000000739.

Adverse events associated with femtosecond laser-assisted cataract surgery reported to the MAUDE database

Affiliations

Adverse events associated with femtosecond laser-assisted cataract surgery reported to the MAUDE database

Lisa R Koenig et al. J Cataract Refract Surg. .

Abstract

Purpose: To report real-world complications associated with femtosecond laser-assisted cataract surgery (FLACS).

Setting: Deidentified database.

Design: Retrospective review.

Methods: The U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database was culled for adverse events involving ophthalmic femtosecond laser during cataract surgery from January 2010 to January 2020. Reports in which the narrative description was inadequate to determine the nature of the complication were excluded. Each report was reviewed for complications, which were subsequently assigned to a complication category as determined by these authors. The number of events per year was determined and compared.

Results: Between January 2010 and January 2020, the MAUDE database search yielded 2927 reports involving femtosecond laser, of which 1927 met inclusion criteria and were included for analysis. These reports revealed 2704 complications: 1115 capsule tears (41.2%), 329 of which required vitrectomy, 372 instances of suction loss, 321 other capsulotomy-related complications, and 432 corneal or incisional complications. Rare complications included 54 reports of miosis, 16 reports of endophthalmitis, and 4 reports of toxic anterior segment syndrome. The number of events in the MAUDE database increased year over year between 2011 and 2017, peaking in 2017 at 303 events.

Conclusions: By drawing on real-world data in the MAUDE database, this study identified the complications of greatest importance in clinical practice. Surgeons should be aware of the spectrum of possible complications to best prepare for all potential outcomes and to effectively counsel patients.

PubMed Disclaimer

References

    1. Nagy Z, Takacs A, Filkorn T, Sarayba M. Initial clinical evaluation of an intraocular femtosecond laser in cataract surgery. J Refract Surg 2009;25:1053–1060
    1. Manning S, Barry P, Henry Y, Rosen P, Stenevi U, Young D, Lundström M. Femtosecond laser-assisted cataract surgery versus standard phacoemulsification cataract surgery: study from the European Registry of Quality Outcomes for Cataract and Refractive Surgery. J Cataract Refract Surg 2016;42:1779–1790
    1. Connell BJ, Kane JX, Vajpayee RB. A comparison of refractive accuracy between conventional and femtosecond laser cataract surgery techniques using modern IOL formulas. Clin Ophthalmol 2021;15:899–907
    1. Day AC, Burr JM, Bennett K, Hunter R, Bunce C, Doré CJ, Nanavaty MA, Balaggan KS, Wilkins MR. Femtosecond laser-assisted cataract surgery compared with phacoemulsification: the FACT non-inferiority RCT. Health Technol Assess 2021;25:1–68
    1. Day AC, Gore DM, Bunce C, Evans JR. Laser-assisted cataract surgery versus standard ultrasound phacoemulsification cataract surgery. Cochrane Database Syst Rev 2016;7:CD010735