Cardiac Output Measurement in Neonates and Children Using Noninvasive Electrical Bioimpedance Compared With Standard Methods: A Systematic Review and Meta-Analysis

Affiliations
Meta-Analysis

Cardiac Output Measurement in Neonates and Children Using Noninvasive Electrical Bioimpedance Compared With Standard Methods: A Systematic Review and Meta-Analysis

Roshni C Mansfield et al. Crit Care Med. .

Abstract

Objective: To systematically review and meta-analyze the validity of electrical bioimpedance-based noninvasive cardiac output monitoring in pediatrics compared with standard methods such as thermodilution and echocardiography.

Data sources: Systematic searches were conducted in MEDLINE and EMBASE (2000-2019).

Study selection: Method-comparison studies of transthoracic electrical velocimetry or whole body electrical bioimpedance versus standard cardiac output monitoring methods in children (0-18 yr old) were included.

Data extraction: Two reviewers independently performed study selection, data extraction, and risk of bias assessment. Mean differences of cardiac output, stroke volume, or cardiac index measurements were pooled using a random-effects model (R Core Team, R Foundation for Statistical Computing, Vienna, Austria, 2019). Bland-Altman statistics assessing agreement between devices and author conclusions about inferiority/noninferiority were extracted.

Data synthesis: Twenty-nine of 649 identified studies were included in the qualitative analysis, and 25 studies in the meta-analyses. No significant difference was found between means of cardiac output, stroke volume, and cardiac index measurements, except in exclusively neonatal/infant studies reporting stroke volume (mean difference, 1.00 mL; 95% CI, 0.23-1.77). Median percentage error in child/adolescent studies approached acceptability (percentage error less than or equal to 30%) for cardiac output in L/min (31%; range, 13-158%) and stroke volume in mL (26%; range, 14-27%), but not in neonatal/infant studies (45%; range, 29-53% and 45%; range, 28-70%, respectively). Twenty of 29 studies concluded that transthoracic electrical velocimetry/whole body electrical bioimpedance was noninferior. Transthoracic electrical velocimetry was considered inferior in six of nine studies with heterogeneous congenital heart disease populations.

Conclusions: The meta-analyses demonstrated no significant difference between means of compared devices (except in neonatal stroke volume studies). The wide range of percentage error reported may be due to heterogeneity of study designs, devices, and populations included. Transthoracic electrical velocimetry/whole body electrical bioimpedance may be acceptable for use in child/adolescent populations, but validity in neonates and congenital heart disease patients remains uncertain. Larger studies in specific clinical contexts with standardized methodologies are required.

PubMed Disclaimer

Conflict of interest statement

Dr. Mansfield is supported by an Imperial Health Charity and Imperial Biomedical Research Centre predoctoral research grant (Grant Reference Number: RFPR2021_29). Imperial Health Charity and Imperial Biomedical Research Centre did not participate in the work. The remaining authors have disclosed that they do not have any potential conflicts of interest.

Comment in

References

    1. Mertens L, Seri I, Marek J, et al.; Writing Group of the American Society of Echocardiography; European Association of Echocardiography; Association for European Pediatric Cardiologists. Targeted Neonatal Echocardiography in the Neonatal Intensive Care Unit: Practice guidelines and recommendations for training. Writing Group of the American Society of Echocardiography (ASE) in collaboration with the European Association of Echocardiography (EAE) and the Association for European Pediatric Cardiologists (AEPC). J Am Soc Echocardiogr. 2011; 24:1057–1078
    1. Singh Y, Gupta S, Groves AM, et al. Expert consensus statement ‘Neonatologist-performed Echocardiography (NoPE)’-training and accreditation in UK. Eur J Pediatr. 2016; 175:281–287
    1. de Boode WP, Singh Y, Gupta S, et al. Recommendations for neonatologist performed echocardiography in Europe: Consensus statement endorsed by European Society for Paediatric Research (ESPR) and European Society for Neonatology (ESN). Pediatr Res. 2016; 80:465–471
    1. Huang SJ, McLean AS. Appreciating the strengths and weaknesses of transthoracic echocardiography in hemodynamic assessments. Cardiol Res Pract. 2012; 2012:894308
    1. de Boode W-P, Osypka M, Soleymani S, Noori S. Chapter 14 - assessment of cardiac output in neonates: Techniques using the Fick principle, indicator dilution technology, Doppler ultrasound, thoracic electrical impedance, and arterial pulse contour analysis. In: Neonatal Questions and Controversies: Hemodynamics and Cardiology. Third Edition. Seri I, Kluckow M (Eds). Philadelphia, PA, Elsevier, 2019, pp 237–263

LinkOut - more resources