. 2021;29(5):823-834.
doi: 10.3233/XST-210898.

Higher agreement in endovascular treatment decision-making than in parametric quantifications among automated CT perfusion software packages in acute ischemic stroke

Affiliations

Higher agreement in endovascular treatment decision-making than in parametric quantifications among automated CT perfusion software packages in acute ischemic stroke

Manuel Pérez-Pelegrí et al. J Xray Sci Technol. 2021.

Abstract

Background and objective: Estimates of parameters used to select patients for endovascular thrombectomy (EVT) for acute ischemic stroke differ among software packages for automated computed tomography (CT) perfusion analysis. To determine impact of these differences in decision making, we analyzed intra-observer and inter-observer agreement in recommendations about whether to perform EVT based on perfusion maps from 4 packages.

Methods: Perfusion CT datasets from 63 consecutive patients with suspected acute ischemic stroke were retrospectively postprocessed with 4 packages of Minerva, RAPID, Olea, and IntelliSpace Portal (ISP). We used Pearson correlation coefficients and Bland-Altman analysis to compare volumes of infarct core, penumbra, and mismatch calculated by Minerva and RAPID. We used kappa analysis to assess agreement among decisions of 3 radiologists about whether to recommend EVT based on maps generated by 4 packages.

Results: We found significant differences between using Minerva and RAPID to estimate penumbra (67.39±41.37mL vs. 78.35±45.38 mL, p < 0.001) and mismatch (48.41±32.03 vs. 61.27±32.73mL, p < 0.001), but not of infarct core (p = 0.230). Pearson correlation coefficients were 0.94 (95%CI:0.90-0.96) for infarct core, 0.87 (95%CI:0.79-0.91) for penumbra, and 0.72 (95%CI:0.57-0.83) for mismatch volumes (p < 0.001). Limits of agreements were (-21.22-25.02) for infarct core volumes, (-54.79-32.88) for penumbra volumes, and (-60.16-34.45) for mismatch volumes. Final agreement for EVT decision-making was substantial between Minerva vs. RAPID (k = 0.722), Minerva vs. Olea (k = 0.761), and RAPID vs. Olea (k = 0.782), but moderate for ISP vs. the other three.

Conclusions: Despite quantitative differences in estimates of infarct core, penumbra, and mismatch using 4 software packages, their impact on radiologists' decisions about EVT is relatively small.

Keywords: Acute ischemic stroke; computed tomography perfusion (CTP); endovascular thrombectomy (EVT); estimates of infarct core; penumbra and mismatch.

PubMed Disclaimer

Similar articles

Cited by

LinkOut - more resources