Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2021 Oct 1;135(4):612-620.
doi: 10.1097/ALN.0000000000003914.

Smart Glasses for Radial Arterial Catheterization in Pediatric Patients: A Randomized Clinical Trial

Affiliations
Randomized Controlled Trial

Smart Glasses for Radial Arterial Catheterization in Pediatric Patients: A Randomized Clinical Trial

Young-Eun Jang et al. Anesthesiology. .

Abstract

Background: Hand-eye coordination and ergonomics are important for the success of delicate ultrasound-guided medical procedures. These can be improved using smart glasses (head-mounted display) by decreasing the head movement on the ultrasound screen. The hypothesis was that the smart glasses could improve the success rate of ultrasound-guided pediatric radial arterial catheterization.

Methods: This prospective, single-blinded, randomized controlled, single-center study enrolled pediatric patients (n = 116, age less than 2 yr) requiring radial artery cannulation during general anesthesia. The participants were randomized into the ultrasound screen group (control) or the smart glasses group. After inducing general anesthesia, ultrasound-guided radial artery catheterization was performed. The primary outcome was the first-attempt success rate. The secondary outcomes included the first-attempt procedure time, the overall complication rate, and operators' ergonomic satisfaction (5-point scale).

Results: In total, 116 children were included in the analysis. The smart glasses group had a higher first-attempt success rate than the control group (87.9% [51/58] vs. 72.4% [42/58]; P = 0.036; odds ratio, 2.78; 95% CI, 1.04 to 7.4; absolute risk reduction, -15.5%; 95% CI, -29.8 to -12.8%). The smart glasses group had a shorter first-attempt procedure time (median, 33 s; interquartile range, 23 to 47 s; range, 10 to 141 s) than the control group (median, 43 s; interquartile range, 31 to 67 s; range, 17 to 248 s; P = 0.007). The overall complication rate was lower in the smart glasses group than in the control group (5.2% [3/58] vs. 29.3% [17/58]; P = 0.001; odds ratio, 0.132; 95% CI, 0.036 to 0.48; absolute risk reduction, 24.1%; 95% CI, 11.1 to 37.2%). The proportion of positive ergonomic satisfaction (4 = good or 5 = best) was higher in the smart glasses group than in the control group (65.5% [38/58] vs. 20.7% [12/58]; P <0.001; odds ratio, 7.3; 95% CI, 3.16 to 16.8; absolute risk reduction, -44.8%; 95% CI, -60.9% to -28.8%).

Conclusions: Smart glasses-assisted ultrasound-guided radial artery catheterization improved the first-attempt success rate and ergonomic satisfaction while reducing the first-attempt procedure time and overall complication rates in small pediatric patients.