Abstract
Introduction: The optimum choice of bearing surfaces in total hip replacement (THR) in the younger and active patient remains controversial. The aim of this study was to report the 10 year clinical outcomes, and a median of 15 year implant survival and incidence of complications in a series of Alumina ceramic-on-ceramic THRs utilising an uncemented shell and cemented stem.
Methods: From January 2004 to December 2007, 175 consecutive patients (195 hips) underwent primary THR. The acetabular components was Trident Peripheral Self Locking (Stryker Orthopaedics) with a third-generation ceramic head and liner (Alumina ceramic, Stryker Orthopaedics). The stem utilised was an Exeter V-40 (Stryker Orthopaedics). Data were collated on demographics, surgical factors, clinical outcomes, radiographic outcomes and revision.
Results: 23 patients (27 THRs) died during the follow-up period at a median of 7.8 (3.8 to 9.0) years post-operatively due to causes unrelated to the THR. Median age at time of surgery was 55 (interquartile range 48-60) years. Median follow-up for surviving patients was 15.2 years. Survivorship for all-cause revision was 97.2%. Increasing patient age at time of surgery was associated with a higher OHS at 10 years (p = 0.022). 32 mm head diameter had an improved OHS at 3 months (p = 0.014) and 10 years (p = 0.030). Posterior surgical approach had a statistically significant better OHS at 3 months (p = 0.015) and 1 year (p < 0.001), but the effect was not significant at 10 years (p = 0.440).
Conclusion: The findings of this study support excellent long-term outcomes and survivorship of Alumina ceramic-on-ceramic bearing in a hybrid THR in a younger population. Surgical factors leading to a more favourable outcome were the use of a 32 mm femoral head and a posterior approach. Increasing age at surgery demonstrated the most sustained improvement in 10 year clinical outcomes.
Keywords: Alumina; Ceramic; Hybrid; Survival; Total hip replacement.
© 2021. The Author(s), under exclusive licence to Springer-Verlag France SAS, part of Springer Nature.
References
-
- National Joint registry Editorial board and contributors. 17th Annual report. 2020 [Cited 1st Oct 2020]. Available from: https://reports.njrcentre.org.uk/Portals/0/PDFdownloads
-
- Kahlenberg CA, Swarup I, Krell EC, Heinz N, Figgie MP (2019) Causes of revision in young patients undergoing total hip arthroplasty. J Arthroplast 34(7):1435–1440. https://doi.org/10.1016/j.arth.2019.03.014 - DOI
-
- Metcalfe D, Peterson N, Wilkinson JM, Perry DC (2018) Temporal trends and survivorship of total hip arthroplasty in very young patients: a study using the National Joint Registry data set. Bone Joint J 100-B(10):1320–1329. https://doi.org/10.1302/0301-620X.100B10.BJJ-2017-1441.R2 - DOI - PubMed
-
- Boyer B, Bordini B, Caputo D, Neri T, Stea S, Toni A (2019) What are the influencing factors on hip and knee arthroplasty survival? Prospective cohort study on 63619 arthroplasties [published correction appears in Orthop Traumatol Surg Res. 2020 Feb;106(1):209]. Orthop Traumatol Surg Res 105(7):1251–1256. https://doi.org/10.1016/j.otsr.2019.07.020
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical