Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Aug 10;11(1):16193.
doi: 10.1038/s41598-021-95607-1.

Clinical validation of optimised RT-LAMP for the diagnosis of SARS-CoV-2 infection

Affiliations

Clinical validation of optimised RT-LAMP for the diagnosis of SARS-CoV-2 infection

Boon Lim et al. Sci Rep. .

Abstract

We have optimised a reverse transcription-loop-mediated isothermal amplification (RT-LAMP) assay for the detection of SARS-CoV-2 from extracted RNA for clinical application. We improved the stability and reliability of the RT-LAMP assay by the addition of a temperature-dependent switch oligonucleotide to reduce self- or off-target amplification. We then developed freeze-dried master mix for single step RT-LAMP reaction, simplifying the operation for end users and improving long-term storage and transportation. The assay can detect as low as 13 copies of SARS-CoV2 RNA per reaction (25-μL). Cross reactivity with other human coronaviruses was not observed. We have applied the new RT-LAMP assay for testing clinical extracted RNA samples extracted from swabs of 72 patients in the UK and 126 samples from Greece and demonstrated the overall sensitivity of 90.2% (95% CI 83.8-94.7%) and specificity of 92.4% (95% CI 83.2-97.5%). Among 115 positive samples which Ct values were less than 34, the RT-LAMP assay was able to detect 110 of them with 95.6% sensitivity. The specificity was 100% when RNA elution used RNase-free water. The outcome of RT-LAMP can be reported by both colorimetric detection and quantifiable fluorescent reading. Objective measures with a digitized reading data flow would allow for the sharing of results for local or national surveillance.

PubMed Disclaimer

Conflict of interest statement

The authors declare no competing interests.

Figures

Figure 1
Figure 1
Stability test of improved RT-LAMP reaction mix. (a) Stability test of RT-LAMP reaction mix containing O117_N and O117_Q. Tubes were added with Synthetic RNA Control 2—MN908947.3 (‘+’) or human genome cDNA (‘−’) followed by heating at 65 °C for the time indicated by T. Each case has three replicates. (b) Dried reaction mixes stored at room temperature up to 3 days. Then they mixed with 20 µL DNase/RNase free water and 5 µL of Synthetic RNA Control 2—MN908947.3 (‘+’) or human genome cDNA (‘−’). The tubes were incubated at 65 °C for 30 min. Top row: Dried reaction mixes containing O117_N; Bottom row: Dried reaction mixes containing O117_Q.
Figure 2
Figure 2
Sensitivity and specificity of improved RT-LAMP assay. (a) Determination of 50% endpoints for O117_Q and O117_N preparations of RT-LAMP assay. Full length transcripts were serially diluted in buffer AVE to achieve the total RNA input/reaction indicated on the y axis. Each dot represents one experimental replicate. Dashed lines indicate 50% endpoint as calculated by the Reed-Muench method. Kits were dried except where indicated for RNA Control 2. Full length transcript indicated on figure. Dried reaction mix containing (b) O117_N and (c) O117_Q were tested against non-SARS human-infective coronaviruses. A and B represent two technical replicates.
Figure 3
Figure 3
Quantitative evaluation of pH-dependent colorimetric RT-LAMP readout. Three quantitative methods (a) Absorbance using 430/560 nm ratio; (b) SYTO 9 fluorescence using 485 nm (excitation) and 500 nm (emission); and (c) Qubit fluorescence using Qubit 2.0 fluorometer were used to assess the RT-LAMP result. (df) are the correlation analysis between RT-qPCR and the three quantitative evaluations, respectively. Each dot represents one experimental replicate. Dotted lines indicate 3× standard deviations above the negative controls.
Figure 4
Figure 4
Detection of SARS-CoV-2 from RNA extracts from swab and saliva specimen. Comparison of RT-LAMP for detecting SARS-CoV-2 RNA in extract vs RT-qPCR Ct value. RT-LAMP results were read by colorimetric changes (see Supplementary Fig. S5).

References

    1. Coronavirus (COVID-19): Getting tested. GOV.UK. https://www.gov.uk/guidance/coronavirus-covid-19-getting-tested. Accessed 21 December 2020.
    1. CDC. Information for Laboratories about Coronavirus (COVID-19). Centers for Disease Control and Prevention. https://www.cdc.gov/coronavirus/2019-ncov/lab/virus-requests.html (2020). Accessed 21 December 2020.
    1. Babiker A, Myers CW, Hill CE, Guarner J. SARS-CoV-2 testing. Am. J. Clin. Pathol. 2020;153:706–708. doi: 10.1093/ajcp/aqaa052. - DOI - PMC - PubMed
    1. Becherer L, et al. Loop-mediated isothermal amplification (LAMP)—Review and classification of methods for sequence-specific detection. Anal. Methods. 2020;12:717–746. doi: 10.1039/C9AY02246E. - DOI
    1. Cui Z, et al. Development of a rapid test kit for SARS-CoV-2: An example of product design. Bio-Des. Manuf. 2020;3:83–86. doi: 10.1007/s42242-020-00075-7. - DOI - PMC - PubMed

Publication types

MeSH terms

Supplementary concepts