Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Jun 21;37(4):1061-1068.
doi: 10.1002/joa3.12585. eCollection 2021 Aug.

Factors affecting signal quality in implantable cardiac monitors with long sensing vector

Affiliations

Factors affecting signal quality in implantable cardiac monitors with long sensing vector

Giovanni B Forleo et al. J Arrhythm. .

Abstract

Purpose: Electrical artefacts are frequent in implantable cardiac monitors (ICMs). We analyzed the subcutaneous electrogram (sECG) provided by an ICM with a long sensing vector and factors potentially affecting its quality.

Methods: Consecutive ICM recipients underwent a follow-up where demographics, body mass index (BMI), implant location, and surface ECG were collected. The sECG was then analyzed in terms of R-wave amplitude and P-wave visibility.

Results: A total of 84 patients (43% female, median age 68 [58-76] years) were enrolled at 3 sites. ICMs were positioned with intermediate inclination (n = 44, 52%), parallel (n = 35, 43%), or perpendicular (n = 5, 6%) to the sternum. The median R-wave amplitude was 1.10 (0.72-1.48) mV with P waves readily visible in 69.2% (95% confidence interval, CI: 57.8%-79.2%), partially visible in 23.1% [95% CI: 14.3%-34.0%], and never visible in 7.7% [95% CI: 2.9%-16.0%] of patients. Men had higher R-wave amplitudes compared to women (1.40 [0.96-1.80] mV vs 1.00 [0.60-1.20] mV, P = .001), while obese people tended to have lower values (0.80 [0.62-1.28] mV vs 1.10 [0.90-1.50] mV, P = .074). The P-wave visibility reached 86.2% [95% CI: 68.3%-96.1%] in patients with high-voltage P waves (≥0.2 mV) at surface ECG. The sECG quality was not affected by implant site.

Conclusion: In ordinary clinical practice, ICMs with long sensing vector provided median R-wave amplitude above 1 mV and reliable P-wave visibility of nearly 70%, regardless of the position of the device. Women and obese patients showed lower but still very good signal quality.

Keywords: P‐wave visibility; R‐wave amplitude; implantable cardiac monitor; implantable loop recorder; long sensing vector.

PubMed Disclaimer

Conflict of interest statement

Daniele Giacopelli is employee of BIOTRONIK Italia. All the other authors have no conflicts relevant to the contents of this study to disclose.

Figures

FIGURE 1
FIGURE 1
Implantation site of the device: the inclination was assessed using the midline of the sternum as reference and defined as perpendicular to sternum (0°≤ angle α ≤20°), with intermediate inclination (20°< angle α ≤60°), or parallel to sternum (60°< angle α ≤90°), while intercostal spaces were used to indicate its cranial/caudal location. Number of patients, median R‐wave amplitude (interquartile range), and proportion of patients with readily visible P waves (95% confidence interval) are reported in the text box for each device inclination
FIGURE 2
FIGURE 2
Example of Implantable cardiac monitor interrogation and sECG (blue signal) measurements: P‐wave visibility was assessed as “readily visible”, ie, visible in all beats (red arrows), while the peak‐to‐peak R‐wave amplitude was measured positioning the automatic dotted vertical lines on 3 different QRS complexes (2 pictured). The amplitude is reported in the red rectangle
FIGURE 3
FIGURE 3
Comparison between the average R‐wave amplitude manually measured on the displayed sECG in supine and standing position and the value automatically provided by the remote monitoring system. P >.05 for each paired comparison with Bonferroni correction
FIGURE 4
FIGURE 4
Subgroup analysis of the median R‐wave amplitude on ICM signal
FIGURE 5
FIGURE 5
P‐wave visibility on ICM signal with supine and standing position
FIGURE 6
FIGURE 6
Subgroup analysis of P‐wave visibility on ICM signal. Abbreviations: ECG: electrocardiogram; ICM: implantable cardiac monitor

References

    1. Brignole M, Vardas P, Hoffman E, Huikuri H, Moya A, Ricci R, et al. EHRA Scientific Documents Committee. Indications for the use of diagnostic implantable and external ECG loop recorders. Europace. 2009;11(5):671–87. - PubMed
    1. Ciconte G, Giacopelli D, Pappone C. The role of implantable cardiac monitors in atrial fibrillation management. J Atr Fibrillation. 2017;10(2):1590. - PMC - PubMed
    1. Ciconte G, Saviano M, Giannelli L, Calovic Z, Baldi M, Ciaccio C, et al. Atrial fibrillation detection using a novel three‐vector cardiac implantable monitor: the atrial fibrillation detect study. Europace. 2017;19(7):1101–8. - PubMed
    1. Bisignani A, De Bonis S, Mancuso L, Ceravolo G, Giacopelli D, Pelargonio G, et al. Are implantable cardiac monitors reliable tools for cardiac arrhythmias detection? An intrapatient comparison with permanent pacemakers. J Electrocardiol. 2020;59:147–50. - PubMed
    1. De Coster M, Demolder A, De Meyer V, Vandenbulcke F, Van Heuverswyn F, De Pooter J. Diagnostic accuracy of R‐wave detection by insertable cardiac monitors. Pacing Clin Electrophysiol. 2020;43(5):511–7. - PubMed

LinkOut - more resources