. 2022 Mar 1;48(3):298-303.
doi: 10.1097/j.jcrs.0000000000000748.

Real-world incidence of monofocal toric intraocular lens repositioning: analysis of the American Academy of Ophthalmology IRIS Registry

Affiliations

Real-world incidence of monofocal toric intraocular lens repositioning: analysis of the American Academy of Ophthalmology IRIS Registry

Brent A Kramer et al. J Cataract Refract Surg. .

Abstract

Purpose: To determine the 12-month incidence of reoperation to realign 2 commercially available types of implanted monofocal toric acrylic intraocular lenses (IOLs).

Setting: American Academy of Ophthalmology IRIS (Intelligent Research in Sight) Registry.

Design: Registry retrospective study.

Methods: Eyes that underwent cataract extraction and were implanted with a TECNIS or AcrySof monofocal toric IOL in 2016 and 2017 were identified. The rate of reoperation for IOL realignment (Current Procedural Terminology code 66825) within 365 days of implantation was determined for each IOL group. Risk factors for repositioning were evaluated using logistic regression modeling.

Results: A total of 6482 eyes were implanted with a monofocal toric IOL, including 2013 (31.06%) with a TECNIS and 4469 (68.94%) with an AcrySof IOL. During the first postoperative year, 87 (1.3%) eyes underwent surgical IOL repositioning. The incidence of repositioning was significantly higher (P < .0001) for TECNIS-implanted (3.1%, 62/2013) than for AcrySof-implanted (0.6%, 25/4469) eyes (odds ratio [OR] 5.6; 95% CI, 3.5-8.9). Younger age (OR 0.76; 95% CI, 0.67-0.86 per 5-year increase) was associated with a higher risk for IOL repositioning.

Conclusions: Real-world analysis of U.S. patients in the IRIS Registry revealed that the rate of surgical IOL repositioning was 5 times higher in eyes implanted with TECNIS than with AcrySof monofocal toric IOLs for astigmatic correction at the time of cataract surgery. These findings should be considered when selecting a toric IOL for correction of astigmatism in cataract patients, particularly in younger patients with a higher risk for misalignment requiring repositioning.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
Incidence of reoperation for IOL realignment ( CPT code 66825) within 365 days of surgery by IOL brand. AcrySof toric IOLs were found to require less additional repositioning than TECNIS toric IOLS (*P < .0001; chi-square test). CPT = Current Procedural Terminology
Figure 2.
Figure 2.
Risk factors for IOL repositioning within 365 days of primary surgery.

References

    1. Hoffmann PC, Hutz WW. Analysis of biometry and prevalence data for corneal astigmatism in 23,239 eyes. J Cataract Refract Surg 2010;36:1479–1485 - PubMed
    1. Day AC, Dhariwal M, Keith MS, Ender F, Perez Vives C, Miglio C, Zou L, Anderson DF. Distribution of preoperative and postoperative astigmatism in a large population of patients undergoing cataract surgery in the UK. Br J Ophthalmol 2019;103:993–1000 - PMC - PubMed
    1. Anderson DF, Dhariwal M, Bouchet C, Keith MS. Global prevalence and economic and humanistic burden of astigmatism in cataract patients: a systematic literature review. Clin Ophthalmol 2018;12:439–452 - PMC - PubMed
    1. U. S. Food and Drug Administration. Premarket approval (PMA). STAAR UV-absorbing silicone posterior chamber IOLs w/toric optic. Available at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P880.... Accessed June 2, 2020
    1. Agresta B, Knorz MC, Donatti C, Jackson D. Visual acuity improvements after implantation of toric intraocular lenses in cataract patients with astigmatism: a systematic review. BMC Ophthalmol 2012;12:41. - PMC - PubMed