Case Reports
. 2021 Jul 29;13(7):e16726.
doi: 10.7759/cureus.16726. eCollection 2021 Jul.

Headache as a Presenting Symptom of Deep Brain Stimulator Generator Failure

Affiliations
Case Reports

Headache as a Presenting Symptom of Deep Brain Stimulator Generator Failure

Trevor Luck et al. Cureus. .

Abstract

While a headache can have a wide variety of clinical presentations, it may occasionally be a red flag for underlying pathology that should prompt further investigation. Here, we present a case report demonstrating headache as an uncommon symptom of deep brain stimulation (DBS) device failure and discuss its clinical significance in the rapidly expanding list of current indications of DBS treatment. A 61-year-old female underwent bilateral hypothalamic DBS implantation for refractory morbid obesity. After a successful course involving significant weight loss, the patient began to experience worsening of her chronic headaches, refractory to her existing regiment. On interrogation, her generator was found to be depleted and its subsequent replacement led to a near total resolution of her headaches. This represents one of the few reported instances of headache as a sign of device failure in DBS treatment, thus adding to the wide possibility of headache presentations and their underlying pathology.

Keywords: dbs; deep brain stimulation; lateral hypothalamic area; mammilotegmental fascisculus; neuromodulation; obesity.

PubMed Disclaimer

Conflict of interest statement

The authors have declared that no competing interests exist.

Figures

Figure 1
Figure 1. (A) Axial postoperative head CT demonstrating bilateral hypothalamic lead placement. (B) Lateral postoperative head x-ray demonstrating bilateral hypothalamic lead placement.

References

    1. Current and future directions of deep brain stimulation for neurological and psychiatric disorders. Lee DJ, Lozano CS, Dallapiazza RF, Lozano AM. J Neurosurg. 2019;131:333–342. - PubMed
    1. Management of deep brain stimulator battery failure: battery estimators, charge density, and importance of clinical symptoms. Fakhar K, Hastings E, Butson CR, Foote KD, Zeilman P, Okun MS. PLoS One. 2013;8:0. - PMC - PubMed
    1. Systematic review of hardware-related complications of deep brain stimulation: do new indications pose an increased risk? Jitkritsadakul O, Bhidayasiri R, Kalia SK, Hodaie M, Lozano AM, Fasano A. Brain Stimul. 2017;10:967–976. - PubMed
    1. Occipital nerve stimulation and deep brain stimulation for refractory cluster headache: a prospective analysis of efficacy over time [PREPRINT] Aibar-Durán JÁ, Holzapfel MJA, Rodríguez RR, Nieto RB, Arnall CR, Teixido JM. J Neurosurg. 2020:1–8. - PubMed
    1. Deep brain stimulation of the hypothalamus leads to increased metabolic rate in refractory obesity. Whiting AC, Sutton EF, Walker CT, et al. World Neurosurg. 2019;121:0–74. - PMC - PubMed

Publication types