Randomized Controlled Trial
. 2021 Dec;64(12):2713-2724.
doi: 10.1007/s00125-021-05555-8. Epub 2021 Sep 8.

Flash glucose monitoring and automated bolus calculation in type 1 diabetes treated with multiple daily insulin injections: a 26 week randomised, controlled, multicentre trial

Affiliations
Randomized Controlled Trial

Flash glucose monitoring and automated bolus calculation in type 1 diabetes treated with multiple daily insulin injections: a 26 week randomised, controlled, multicentre trial

Anna Lilja Secher et al. Diabetologia. 2021 Dec.

Erratum in

Abstract

Aims/hypothesis: We aimed to compare the effects of intermittently scanned continuous glucose monitoring (isCGM) and carbohydrate counting with automated bolus calculation (ABC) with usual care.

Methods: In a randomised, controlled, open-label trial carried out at five diabetes clinics in the Capital Region of Denmark, 170 adults with type 1 diabetes for ≥1 year, multiple daily insulin injections and HbA1c > 53 mmol/mol (7.0%) were randomly assigned 1:1:1:1 with centrally prepared envelopes to usual care (n = 42), ABC (n = 41), isCGM (n = 48) or ABC+isCGM (n = 39). Blinded continuous glucose monitoring data, HbA1c and patient-reported outcomes were recorded at baseline and after 26 weeks. The primary outcome was change in time in range using isCGM vs usual care.

Results: Baseline characteristics were comparable across arms: mean age 47 (SD 13.7) years, median (IQR) diabetes duration 18 (10-28) years and HbA1c 65 (61-72) mmol/mol (8.1% [7.7-8.7%]). Change in time in range using isCGM was comparable to usual care (% difference of 3.9 [-12-23], p = 0.660). The same was true for the ABC and ABC+isCGM arms and for hypo- and hyperglycaemia. Also compared with usual care, using ABC+isCGM reduced HbA1c (4 [95% CI 1, 8] mmol/mol) (0.4 [0.1, 0.7] %-point) and glucose CV (11% [4%, 17%]) and improved treatment satisfaction, psychosocial self-efficacy and present life quality. Treatment satisfaction also improved by using isCGM alone vs usual care. Statistical significance was maintained after multiple testing adjustment concerning glucose CV and treatment satisfaction with ABC+isCGM, and treatment satisfaction with isCGM. Discontinuation was most common among ABC only users, and among completers the ABC was used 4 (2-5) times/day and the number of daily isCGM scans was 5 (1-7) at study end.

Conclusions/interpretation: isCGM alone did not improve time in range, but treatment satisfaction increased in technology-naive people with type 1 diabetes and suboptimal HbA1c. The combination of ABC+isCGM appears advantageous regarding glycaemic variables and patient-reported outcomes, but many showed resistance towards ABC.

Trial registration: ClinicalTrials.gov NCT03682237.

Funding: The study is investigator initiated and financed by the Capital Region of Denmark.

Keywords: Automated bolus calculation; Continuous glucose monitoring; Flash glucose monitoring; Glycaemic control; HbA1c; Multiple daily insulin injections; Time in range; Type 1 diabetes.

PubMed Disclaimer

References

    1. Beck RW, Bergenstal RM, Laffel LM, Pickup JC (2019) Advances in technology for management of type 1 diabetes. Lancet 394(10205):1265–1273. https://doi.org/10.1016/S0140-6736(19)31142-0 - DOI - PubMed
    1. Beck RW, Riddlesworth T, Ruedy K et al (2017) Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections. JAMA 317(4):371–378. https://doi.org/10.1001/jama.2016.19975
    1. Lind M, Polonsky W, Hirsch IB et al (2017) Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections. JAMA 317(4):379–387. https://doi.org/10.1001/jama.2016.19976
    1. American Diabetes Association (2021) Diabetes technology: standards of medical care in diabetes−2021. Diabetes Care 44:S85–S99. https://doi.org/10.2337/dc21-S007 - DOI
    1. American Diabetes Association (2021) Glycemic targets: standards of medical care in diabetes−2021. Diabetes Care 44:S73–S84. https://doi.org/10.2337/dc21-S006 - DOI