Better late than never: correcting the error in the Exhalyzer nitrogen washout system

Affiliations
Editorial

Better late than never: correcting the error in the Exhalyzer nitrogen washout system

Alexander Horsley et al. J Appl Physiol (1985). .
No abstract available

Keywords: Exhalyzer; cystic fibrosis; lung function; nitrogen washout.

PubMed Disclaimer

Conflict of interest statement

A.H. has grants from NIHR, CF Trust, and CF Foundation. A.H. has worked on multiple-breath washout using the Innocor system and has collaborated with Innovision ApS, the manufacturer of that device.

Comment on

References

    1. Ratjen F, Hug C, Marigowda G, Tian S, Huang X, Stanojevic S, Milla CE, Robinson PD, Waltz D, Davies JC; VX14-809-109 Investigator Group. Efficacy and safety of lumacaftor and ivacaftor in patients aged 6–11 years with cystic fibrosis homozygous for F508del-CFTR: a randomised, placebo-controlled phase 3 trial. Lancet Respir Med 5: 557–567, 2017. [Erratum in Lancet Respir Med 5: e28, 2017]. doi:10.1016/S2213-2600(17)30215-1. - DOI - PubMed
    1. Subbarao P, Milla C, Aurora P, Davies JC, Davis SD, Hall GL, Heltshe S, Latzin P, Lindblad A, Pittman JE, Robinson PD, Rosenfeld M, Singer F, Starner TD, Ratjen F, Morgan W. Multiple-breath washout as a lung function test in cystic fibrosis. A Cystic Fibrosis Foundation Workshop report. Ann Am Thorac Soc 12: 932–939, 2015. [Erratum in Ann Am Thorac Soc 14: 145, 2017]. doi:10.1513/AnnalsATS.201501-021FR. - DOI - PMC - PubMed
    1. Sandvik RM, Gustafsson PM, Lindblad A, Robinson PD, Nielsen KG. Improved agreement between N2 and SF6 multiple-breath washout in healthy infants and toddlers with improved EXHALYZER D sensor performance. J Appl Physiol (1985) 131: 107–118, 2021. doi:10.1152/japplphysiol.00129.2021. - DOI - PubMed
    1. Wyler F, Oestreich MA, Frauchiger BS, Ramsey KA, Latzin P. Correction of sensor crosstalk error in Exhalyzer D multiple-breath washout device significantly impacts outcomes in children with cystic fibrosis. J Appl Physiol (1985) 131: 1148–1156, 2021.doi:10.1152/japplphysiol.00338.2021. - DOI - PubMed
    1. Jensen R, Stanojevic S, Gibney K, Salazar JG, Gustafsson P, Subbarao P, Ratjen F. Multiple breath nitrogen washout: a feasible alternative to mass spectrometry. PLoS One 8: e56868, 2013. doi:10.1371/journal.pone.0056868. - DOI - PMC - PubMed

LinkOut - more resources