Knotted Transosseous-Equivalent Technique for Rotator Cuff Repair Shows Superior Biomechanical Properties Compared With a Knotless Technique: A Systematic Review and Meta-analysis
- PMID: 34606935
- DOI: 10.1016/j.arthro.2021.09.017
Knotted Transosseous-Equivalent Technique for Rotator Cuff Repair Shows Superior Biomechanical Properties Compared With a Knotless Technique: A Systematic Review and Meta-analysis
Abstract
Purpose: To compare the biomechanical properties of the knotted versus knotless transosseous-equivalent (TOE) techniques for rotator cuff repair (RCR).
Methods: A systematic review was performed according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines using PubMed, Embase, and the Cochrane Library to identify studies that compared the biomechanical properties of knotted and knotless TOE RCR techniques. The search phrase used was as follows: (Double Row) AND (rotator cuff) AND (repair) AND (biomechanical). Evaluated properties included ultimate load to failure, cyclic displacement, stiffness, footprint characteristics, and failure mode.
Results: Eight studies met the inclusion criteria, including a total of 67 specimens in each group. Of 6 studies reporting on ultimate load to failure, 4 found tendons repaired with the knotted TOE technique to experience significantly higher ultimate load to failure compared with knotless TOE repairs (knotted range, 323.5-549.0 N; knotless range, 166.0-416.8 N; P < .05). Of 6 studies reporting on failure stiffness, 2 found knotted TOE repairs to have significantly higher failure stiffness compared with knotless TOE repairs (knotted range, 30.0-241.8 N/mm; knotless range, 28.0-182.5 N/mm; P < .05), whereas 1 study found significantly higher failure stiffness in knotless TOE repairs compared with knotted TOE repairs (P = .039). Cyclic gap formation favored the knotted TOE group in 2 of 3 studies (knotted range, 0.6-5.2 mm; knotless range, 0.4-9.1 mm; P < .05). The most common mode of failure in both groups was suture tendon tear.
Conclusions: On the basis of the included cadaveric studies, rotator cuff tendons repaired via the knotted TOE technique display superior time-zero biomechanical properties, including greater ultimate load to failure, compared with rotator cuffs repaired via the knotless TOE technique. Suture tearing through the tendon remains a common failure method for both techniques.
Clinical relevance: The results of this systematic review provide helpful insight into the biomechanical differences between 2 popular techniques for RCR. Although these results should be carefully considered by surgeons who are using either of these techniques in the operating room, they should not be mistaken for direct clinical applicability because cadaveric studies may not directly correlate to clinical outcomes.
Copyright © 2021 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Leptin Enhances M1 Macrophage Polarization and Impairs Tendon-Bone Healing in Rotator Cuff Repair: A Rat Model.Clin Orthop Relat Res. 2025 May 1;483(5):939-951. doi: 10.1097/CORR.0000000000003428. Epub 2025 Feb 19. Clin Orthop Relat Res. 2025. PMID: 39982019
-
Rotator Cuff Repair Augmentation by Direct Interlocking of an Overlayed Nonwoven Polyethylene Terephthalate Patch Substantially Increases Repair Strength: A Biomechanical Comparison in Ovine Shoulders.Orthop J Sports Med. 2025 Jul 21;13(7):23259671251356629. doi: 10.1177/23259671251356629. eCollection 2025 Jul. Orthop J Sports Med. 2025. PMID: 40697811 Free PMC article.
-
Transosseous-equivalent rotator cuff repair: a systematic review on the biomechanical importance of tying the medial row.Arthroscopy. 2013 Feb;29(2):377-86. doi: 10.1016/j.arthro.2012.11.008. Arthroscopy. 2013. PMID: 23369482
-
Smaller Gap Formation With Suture Anchor Fixation Than Traditional Transpatellar Sutures in Patella and Quadriceps Tendon Rupture: A Systematic Review.Arthroscopy. 2022 Jul;38(7):2321-2330. doi: 10.1016/j.arthro.2022.01.012. Epub 2022 Jan 20. Arthroscopy. 2022. PMID: 35066110
-
Comparison of clinical and biomechanical results of arthroscopic rotator cuff repair using conventional and triple-row suture-bridge techniques.J Orthop Surg (Hong Kong). 2025 May-Aug;33(2):10225536251364198. doi: 10.1177/10225536251364198. Epub 2025 Jul 30. J Orthop Surg (Hong Kong). 2025. PMID: 40735888
Cited by
-
Comparison of Knotless and Knotted Single-Anchor Repair for Ruptures of the Upper Subscapularis Tendon: Outcomes at 2-Year Follow-up.Orthop J Sports Med. 2022 Mar 15;10(3):23259671221083591. doi: 10.1177/23259671221083591. eCollection 2022 Mar. Orthop J Sports Med. 2022. PMID: 35309232 Free PMC article.
-
The "Greenhouse" Technique Using Knotless Single-Row Suture Bridge Combined With Bone Marrow Stimulation for the Arthroscopic Treatment of Rotator Cuff Tears.Arthrosc Tech. 2022 Jan 20;11(2):e189-e196. doi: 10.1016/j.eats.2021.10.010. eCollection 2022 Feb. Arthrosc Tech. 2022. PMID: 35155112 Free PMC article.
-
A comparison between knotted and knotless medial row of suture bridge technique in arthroscopic rotator cuff repair surgery: a meta-analysis.J Orthop Surg Res. 2023 May 8;18(1):338. doi: 10.1186/s13018-023-03812-7. J Orthop Surg Res. 2023. PMID: 37158896 Free PMC article. Review.
-
Modified Double-Pulley And Rip-Stop Suture Bridge Technique In Arthroscopic Rotator Cuff Repair.Arthrosc Tech. 2023 Jan 18;12(2):e301-e305. doi: 10.1016/j.eats.2022.11.007. eCollection 2023 Feb. Arthrosc Tech. 2023. PMID: 36879864 Free PMC article.
-
Blended Suture-bridge Technique for Arthroscopic Rotator Cuff Repair.Arthrosc Tech. 2023 Mar 27;12(4):e569-e574. doi: 10.1016/j.eats.2022.12.011. eCollection 2023 Apr. Arthrosc Tech. 2023. PMID: 37138682 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous