. 2022 May 4;7(3):737-746.
doi: 10.1093/jalm/jfab112.

Overcome Isomer Interference in 1α,25-Dihydroxyvitamin D Quantitation by Liquid Chromatography-Tandem Mass Spectrometry

Affiliations

Overcome Isomer Interference in 1α,25-Dihydroxyvitamin D Quantitation by Liquid Chromatography-Tandem Mass Spectrometry

Zhicheng Jin et al. J Appl Lab Med. .

Abstract

Background: The circulating concentration of 1α,25-dihydroxyvitamin D [1α,25(OH)2D] is very low, and the presence of multiple isomers may lead to inaccurate quantitation if not separated prior to analysis. Antibody-based immunoextraction procedures are sometimes used to remove structurally related isomers of 1α,25(OH)2D prior to an LC-MS/MS analysis. However, immunoextraction increases sample preparation time and cost. In addition, some dihydroxyvitamin D metabolites are not completely removed by immunoextraction.

Method: We developed an HPLC method using a phenyl-hexyl column to investigate interfering isomers of 1α,25(OH)2D.

Result: Using this method, 4-phenyl-1,2,4-triazoline-3,5-dione (PTAD) derivatization product of 1α,25(OH)2D was found to be present as 2 epimers, which were separated chromatographically with an area ratio of 2:1. PTAD derivatized metabolite of 25-hydroxyvitamin D3 [i.e., 4β,25-dihydroxyvitamin D3 (4β,25(OH)2D3)] eluted out between 6R and 6S epimers of derivatized 1α,25(OH)2D3. If not chromatographically resolved, 4β,25(OH)2D can affect 1α,25(OH)2D quantitation. In a method comparison study, it was found that the presence of 4β,25(OH)2D produced positive bias up to 127% on 1α,25(OH)2D3 quantitation.

Conclusion: The LC-MS/MS method we developed without an immunoextraction procedure was able to resolve the major interference peak from 1α,25(OH)2D and achieved reliable quantitation of 1α,25(OH)2D.

Keywords: 1,25-dihydroxyvitamin D; 4β,25-dihydroxyvitamin D; LC-MS/MS; epimer; isomer interference.