. 2021 Nov;33(11):E857-E862.
doi: 10.25270/jic/20.00655. Epub 2021 Oct 15.

A Comparison Between Gore Cardioform and Amplatzer Septal Occluder for Percutaneous Closure of Patent Foramen Ovale Associated With Atrial Septal Aneurysm: Clinical and Echocardiographic Outcomes

Affiliations
Free article

A Comparison Between Gore Cardioform and Amplatzer Septal Occluder for Percutaneous Closure of Patent Foramen Ovale Associated With Atrial Septal Aneurysm: Clinical and Echocardiographic Outcomes

Carmine Musto et al. J Invasive Cardiol. 2021 Nov.
Free article

Abstract

Objectives: To evaluate the short- and long-term clinical and echocardiographic outcomes of the percutaneous closure of the patent foramen ovale (PFO) with an atrial septal aneurysm (ASA) using 2 different devices.

Methods: We enrolled 100 patients with PFO and ASA. Fifty consecutive patients had transcatheter closure of the PFO with the Gore Cardioform septal occluder (GSO) (Gore Medical) and a second group of 50 consecutive patients with the Amplatzer PFO occluder (APO) (Abbott). A clinical and transthoracic echocardiographic follow-up was performed at 1, 3, 6, and 12 months after the implant procedure. The primary endpoint was the incidence of moderate-to-severe residual right-to-left shunting (rRLS) at the 6-month follow-up. The procedural results and the recurrence of embolic events at 1 year were also investigated.

Results: The procedure was successful in all patients. The immediate postprocedural moderate-to-severe rRLS incidence was similar between the 2 groups (GSO 14% vs APO 12%; P=NS) as well as the incidence of moderate-to-severe rRLS at the 6-month follow-up (GSO 4% vs APO 4%; P=non-significant). In only 1 patient of the GSO group, there was a persistent moderate rRLS at the 1-year follow-up. The 6-month and 1-year complete occlusion rate for all subjects was 93% and 96%, respectively. No devices embolized and no death or recurrent embolic events were observed during hospitalization through the 1-year follow-up.

Conclusions: GSO and APO devices appear to be safe and effective devices for the percutaneous closure of a PFO with ASA, showing similar results for the presence of rRLS at the 6-month follow-up, complete occlusion rate, and clinical embolic recurrences.

Keywords: Amplatzer septal occluder; Gore Cardioform septal occluder; atrial septal aneurysm; intracardiac echography; patent foramen ovale; residual right-to-left shunting.