Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2022 Feb;38(2):333-341.
doi: 10.1007/s00381-021-05390-7. Epub 2021 Oct 15.

High-resistance proximal "scaled" ventricular catheters

Affiliations

High-resistance proximal "scaled" ventricular catheters

David Qi et al. Childs Nerv Syst. 2022 Feb.

Abstract

Purpose: Prove the concept of high-resistance proximal catheters for valve-independent treatment of hydrocephalus.

Methods: A preliminary design process yielded optimal high-resistance proximal ventricular catheters with a "scaled" design and parallel-oriented, U-shaped inlets. Prototypes were manually constructed using carving tools to stamp through silicone tubings. A testing apparatus was developed to simulate cerebrospinal fluid flow through a catheter, and the prototypes were tested against a control catheter for exhibition of an "on/off" phenomenon whereby no flow occurs at low pressures, and flow begins beyond a pressure threshold. Flow distribution was visualized with India ink. Regression analysis was performed to determine linearity.

Results: The new designs showed varying amounts of improved flow control with the "scaled" design showing the most practical flow rate control across various pressures, compared to the standard catheter; however, no true "on/off" phenomenon was observed. The "scaled" design showed various degrees of dynamism; its flow rate can be time dependent, and certain maneuvers such as flushing and bending increased flow rate temporarily. Variation in the number of inlets within each "scaled" prototype also affected flow rate. Contrastingly, the flow rate of standard catheters was found to be independent of the number of inlet holes. Ink flow showed even flow distribution in "scaled" prototypes.

Conclusions: This initial feasibility study showed that high-resistance ventricular catheters can be designed to mimic the current/valved system. The "scaled" design demonstrated the best flow control, and its unique features were characterized.

Keywords: Cerebrospinal fluid; Flow rate; Hydrocephalus; Shunt.

PubMed Disclaimer

References

    1. Galarza M, Etus V, Sosa F, Argañaraz R, Mantese B, Gazzeri R, Montoya CG, de la Rosa P, Guerrero AL, Chaban G, Giménez A, Amigó JM (2020) Flow ventricular catheters for shunted hydrocephalus: initial clinical results. Childs Nerv Syst 37:903–911. https://doi.org/10.1007/s00381-020-04941-8 - DOI - PubMed
    1. Galarza M, Giménez A, Amigó JM, Schuhmann M, Gazzeri R, Thomale U, McAllister JP (2018) Next generation of ventricular catheters for hydrocephalus based on parametric designs. Childs Nerv Syst 34:267–276. https://doi.org/10.1007/s00381-017-3565-0 - DOI - PubMed
    1. Galarza M, Giménez A, Valero J, Pellicer OP, Amigó JM (2014) Computational fluid dynamics of ventricular catheters used for the treatment of hydrocephalus: a 3D analysis. Childs Nerv Syst 30:105–116. https://doi.org/10.1007/s00381-013-2226-1 - DOI - PubMed
    1. Giménez Á, Galarza M, Pellicer O, Valero J, Amigó JM (2016) Influence of the hole geometry on the flow distribution in ventricular catheters for hydrocephalus. Biomed Eng Online 15:71. https://doi.org/10.1186/s12938-016-0182-1 - DOI - PubMed - PMC
    1. Harris CA, McAllister JP (2011) Does drainage hole size influence adhesion on ventricular catheters? Childs Nerv Syst 27:1221–1232. https://doi.org/10.1007/s00381-011-1430-0 - DOI - PubMed

LinkOut - more resources