Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Case Reports
. 2021 Sep;53(3):193-198.
doi: 10.1182/ject-2100017.

A Novel Method to Safely De-Air a HeartWare System in a Single-Ventricle Patient by Utilizing ECMO and a Minimized CPB Circuit

Affiliations
Case Reports

A Novel Method to Safely De-Air a HeartWare System in a Single-Ventricle Patient by Utilizing ECMO and a Minimized CPB Circuit

Tiffany M Robb et al. J Extra Corpor Technol. 2021 Sep.

Abstract

The survival of congenital heart disease (CHD) patients with single-ventricle (SV) physiology has markedly increased as a result of advances in operative techniques and postsurgical management. Nonetheless, these patients remain highly susceptible to end-stage heart failure requiring cardiac replacement therapies at early ages. Given a worldwide shortage of transplantable organs, mechanical circulatory support (MCS) represents an alternative treatment option. The significant heterogeneity of the SV population presents unique indications for MCS that have begun to be evaluated. This case study describes a 12-year-old female with heterotaxy syndrome and an SV condition, previously palliated with a Fontan operation at another institution. The patient was placed on veno-arterial (VA) extracorporeal membrane oxygenation (ECMO) during prolonged cardiopulmonary resuscitation, and later underwent HeartWare ventricular assist device (HVAD) implantation as a bridge to transplantation (BTT). A novel method was chosen to optimize careful de-airing of the heart through a minimized cardiopulmonary bypass (CPB) setup, during full ECMO support and surgical insertion of the HeartWare. The ascending aorta was vented proximal to the HVAD outflow graft anastomosis through a minimized CPB circuit at <10% of the ECMO flow rate. This circuit adaption allowed for euvolemic resuscitation via connection from the minimized CPB circuit to the venous limb of the ECMO circuit. The transition from VA-ECMO to the HeartWare was well tolerated despite a challenging sternotomy and cardiac anomaly. A minimized bypass circuit proved efficacious for the benefit of volume resuscitation and safe de-airing of the HVAD while on ECMO support. The literature is limited concerning safe practices for implantation of durable VADs in complex SV patients coupled with those transitioning from varying modalities of MCS. As SV survivability regresses to heart failure, it is essential that we share techniques that aim to improve the long-term outcomes for successful BTT or bridge to decision (BTD).

Keywords: HeartWare; cardiopulmonary bypass; congenital heart disease; extracorporeal membrane oxygenation; heart failure; left ventricular assist device; single ventricle.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
Diagram of a minimized CPB circuit setup for the safe de-airing of the heart following HeartWare HVAD insertion, and reinfusion of vented blood from the aortic root while supported on VA-ECMO. (A) HVAD system. (B) Minimized CPB circuit for venting of the aortic root to the CPB reservoir. (C) Volume resuscitation circuit for reinfusion of filtered and oxygenated blood from the CPB setup directly to the venous limb of the ECMO circuit.

Similar articles

Cited by

References

    1. Miller JR, Lancaster TS, Callahan C, et al. . An overview of mechanical circulatory support in single-ventricle patients. Transl Pediatr. 2018;7:151–61. - PMC - PubMed
    1. Weinstein S, Bello R, Pizarro C, et al. . The use of the Berlin Heart EXCOR in patients with functional single ventricle. J Thorac Cardiovasc Surg. 2014;147:697–704, discussion 704–5. - PubMed
    1. Morales D, Almond CS, Jaquiss RD, et al. . Bridging children of all sizes to cardiac transplantation: The initial multi-center North American experience with Berlin Heart EXCOR ventricular assist device. J Heart Lung Transplant. 2011;30:1–8. - PubMed
    1. Gewillig M, Goldberg DJ.. Failure of the Fontan circulation. Heart Fail Clin. 2014;10:105–16. - PubMed
    1. Griselli M, Sinha R, Jang S, et al. . Mechanical circulatory support for single ventricle failure. Front Cardiovasc Med. 2018;5:115. - PMC - PubMed

Publication types