. 2022 Oct;77(5):832-836.
doi: 10.1080/17843286.2021.1992583. Epub 2021 Oct 19.

Asparaginase-induced pseudohyponatremia, a case-driven working strategy in pediatric patients

Affiliations

Asparaginase-induced pseudohyponatremia, a case-driven working strategy in pediatric patients

A Evenepoel et al. Acta Clin Belg. 2022 Oct.

Abstract

Objectives: When using indirect ion-selective electrode (ISE) methods, hypertriglyceridemia leads to pseudohyponatremia due to water displacement artifacts. Multiple strategies exist to minimize this interference. Our objective was to create a patient-friendly one-tube-fits-all testing setup without compromising the method robustness.

Methods: Four strategies were evaluated in a single patient with hypertriglyceridemia. Additionally, the interchangeability between the Cobas 8000 and ABL Flex was evaluated on samples (n = 2274) with different total protein (TP) concentrations. Finally, a proof-of-concept (n = 40) was performed by re-measuring the routine sample with the ABL90 Flex.

Results: ABL90 flex results and calculated sodium did not suffer from the presence of high triglyceride levels. We did not observe any significant differences between the three groups (P > 0.05) of sample types (arterial vs. venous plasma vs. venous whole blood after mixing up) nor for the analysers (Roche vs. ABL90 Flex). Passing-Bablok and Bland-Altman tests revealed interchangeability.

Conclusion: In future cases of hypertriglyceridemia, 1500 mg/dL will be used as a preliminary threshold for reliable sodium determination. Routine Li-heparin samples can be used for accurate sodium determination without any need for extra arterial or venous blood gas tubes, offering a patient-friendly test setup for similar cases.

Keywords: Pseudohyponatremia; asparaginase; hypertriglyceridemia; ion-selective electrode; pediatric.