Abstract
Objectives: The aim of this retrospective study was to determine if Cormatrix® (CM) represents a safe alternative to conventional patch materials used in congenital heart surgery.
Methods: A total of 57 paediatric patients who underwent cardiac surgery using an Extracellular Matrix Bioscaffold (CM) were categorized into 4 groups according to the patch implant location. Patch-related complications and reintervention rates were analysed. A subgroup of 18 patients was subsequently compared to a matched group of 36 patients who underwent similar surgical procedures with autologous pericardium as patch material.
Results: No patient died during hospitalization. There were 2 late deaths, not related to the implanted CM patch. Fourteen (66.7%) out of 21 patients with arterial patch plasty developed progressive vessel/right ventricular outflow tract stenosis or aneurysm. All 3 patients with a valved CM conduit developed haemodynamically relevant valve stenosis or regurgitation. A total of 18 (31.5%) patients needed reintervention and 12 (21.1%) related to CM. Four (7%) patients needed surgical treatment with operative removal of the stenosis. Redo valve replacement was performed on 2 (3.5%) patients. Six (10.5%) patients required an interventional cardiology procedure at a median interval of 5 months from surgery. The subgroup analysis revealed a significantly lower patch-related reintervention rate in patients treated with autologous pericardium when compared to CM (P = 0.006).
Conclusions: CM is safe for atrial and ventricular defect closure. The use of CM for arterial vessel reconstruction is associated with higher reintervention rates when compared to autologous pericardium.
Keywords: Congenital heart defect Bioscaffold; Cormatrix®; Small intestinal submucosa-extracellular matrix.
© The Author(s) 2021. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery.
References
-
- Dolk H, Loane M, Garne E; European Surveillance of Congenital Anomalies (EUROCAT) Working Group. Congenital heart defects in Europe: prevalence and perinatal mortality, 2000 to 2005. Circulation 2011;123:841–9. - PubMed
-
- Van Der Linde D, Konings EE, Slager MA, Witsenburg M, Helbing WA, Takkenberg JJ. et al. Birth prevalence of congenital heart disease worldwide: a systematic review and meta-analysis. J Am Coll Cardiol 2011;58:2241–7. - PubMed
-
- Witt RG, Raff G, Van Gundy J, Rodgers-Ohlau M, Si MS.. Short-term experience of porcine small intestinal submucosa patches in paediatric cardiovascular surgery. Eur J Cardiothorac Surg 2013;44:72–6. - PubMed
-
- Padalino MA, Quarti A, Angeli E, Frigo AC, Vida VL, Pozzi M. et al. Early and mid-term clinical experience with extracellular matrix scaffold for congenital cardiac and vascular reconstructive surgery: a multicentric Italian study. Interact CardioVasc Thorac Surg 2015;21:40–9. - PubMed
-
- Holubec T, Caliskan E, Sündermann SH, Starck CT, Plass A, Bettex D. et al. The use of extracellular matrix patches in cardiac surgery. J Card Surg 2015;30:145–8. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical