Hypothermic oxygenated machine perfusion for extended criteria donor allografts: Preliminary experience with extended organ preservation times in the setting of organ reallocation
- PMID: 34724239
- DOI: 10.1111/aor.14103
Hypothermic oxygenated machine perfusion for extended criteria donor allografts: Preliminary experience with extended organ preservation times in the setting of organ reallocation
Abstract
Background: In times of critical organ shortage, poor organ pool utilization and increased use of extended-criteria donor (ECD) allografts remain a major problem. Hypothermic oxygenated machine perfusion (HOPE) has emerged as a promising and feasible strategy in ECD liver transplantation (LT). However, potential safety limits regarding the duration of perfusion are yet to be explored. Besides marginal allograft quality (steatosis), prolonged cold ischemia time remains the most important factor for a high number of liver allografts being declined for transplantation.
Patients and methods: Two ECD-allografts were each allocated to two recipients, who proved to be unsuitable to receive the assigned allograft upon arrival at the transplant center. The organs were reallocated by Eurotransplant and accepted by our center for two different backup patients. During that time, HOPE was commenced and continued until the recipient hepatectomy was completed. Postoperative allograft function was assessed by serum levels of alanine aminotransferase (ALT), aspartate aminotransferase (AST), bilirubin, and International Normalized Ratio. Incidence of early allograft dysfunction (EAD), postoperative complications, and length of hospital stay were analyzed.
Results: HOPE was applied for 4 h 35 min and 4 h 20 min, resulting in a total cold preservation time of 17 h 29 min and 15 h 20 min, respectively. Both recipients displayed decreasing serum transaminases and bilirubin levels postoperatively. No EAD or major postoperative complications occurred in either patient. Serum ALT and AST levels were within the normal range at discharge.
Conclusions: Extended HOPE enables the safe extension of preservation time for up to 18 h in human LT. End-ischemic HOPE may significantly improve organ pool utilization, while simultaneously facilitating operating room logistics and preventing organ injury.
Keywords: donation after brain death; extended criteria donor; extended preservation time; hypothermic oxygenated machine perfusion; liver transplantation; operating room logistics.
© 2021 The Authors. Artificial Organs published by International Center for Artificial Organ and Transplantation (ICAOT) and Wiley Periodicals LLC.
References
REFERENCES
-
- Czigany Z, Lurje I, Schmelzle M, Schöning W, Öllinger R, Raschzok N, et al. Ischemia-reperfusion injury in marginal liver grafts and the role of hypothermic machine perfusion: molecular mechanisms and clinical implications. J Clin Med. 2020;9:846.
-
- Czigany Z, Lurje I, Tolba RH, Neumann UP, Tacke F, Lurje G. Machine perfusion for liver transplantation in the era of marginal organs-new kids on the block. Liver Int. 2019;39:228-49.
-
- de Meijer VE, Fujiyoshi M, Porte RJ. Ex situ machine perfusion strategies in liver transplantation. J Hepatol. 2019;70:203-5.
-
- Dutkowski P, Polak WG, Muiesan P, Schlegel A, Verhoeven CJ, Scalera I, et al. First comparison of hypothermic oxygenated perfusion versus static cold storage of human donation after cardiac death liver transplants: an international-matched case analysis. Ann Surg. 2015;262:764-71; discussion 70-1.
-
- Guarrera JV, Henry SD, Samstein B, Odeh-Ramadan R, Kinkhabwala M, Goldstein MJ, et al. Hypothermic machine preservation in human liver transplantation: the first clinical series. Am J Transplant. 2010;10:372-81.
MeSH terms
Substances
LinkOut - more resources