Sweeping-beam technique with electrons for large treatment areas as total skin irradiation : Dosimetric and technical aspects of a modified Stanford technique
- PMID: 34729625
- DOI: 10.1007/s00066-021-01859-1
Sweeping-beam technique with electrons for large treatment areas as total skin irradiation : Dosimetric and technical aspects of a modified Stanford technique
Abstract
Purpose: Total skin electron beam therapy (TSEBT) is still a technical and therapeutic challenge today. Thus, we developed TSEBT using a sweeping-beam technique.
Methods: For treatment delivery, a linear accelerator Versa HD (ELEKTA, Stockholm, Sweden) with high-dose-rate electrons (HDRE) was used with a dose rate of 9000 MU/min. Dosimetry quality assurance was performed by multiple measurements with film dosimetry, 2D array, and Roos chamber.
Results: Clinical experience shows that treatment durations of 75 to 90 min are usual for the Stanford technique without using HDRE. With this new sweeping-beam irradiation technique, the total treatment time of a daily fraction could be reduced to 20 min while keeping over- and underdosing low. The treatment area is about 60 cm × 200 cm and the dose distribution is uniform within 2% and 5% in vertical and horizontal directions, respectively. Initially, the electron energy of 6 MeV is reduced to 3.2 MeV by 1‑cm polymethylmethacrylat (PMMA) scatter and the irradiation conditions of a source-surface distance (SSD) of 350 cm. The photon contamination drops to under 1%.
Conclusion: These results show that the mean dose to total skin varies between 1.3 and 1.8 Gy. The sweeping-beam technique with electrons has a homogeneous dose distribution in connection with a short treatment time.
Keywords: High-dose-rate electrons; In situ film dosimetry; Mycosis fungoides; Total skin electron beam therapy (TSEBT); Total skin electron irradiation (TSEI).
© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany.
References
-
- Asbell SO, Siu J, Lightfoot DA, Brady LW (1980) Individualized eye shields for use in electron beam therapy as well as low-energy photon irradiation. Int J Radiat Oncol Biol Phys 6(4):519–521 - DOI
-
- Bin Xu L (2009) Commissioning of a GafChromic EBT film dosimetry protocol at ionizing radiation standards group of National Research Council
-
- DIN 6800‑2. Dosismessverfahren nach der Sondenmethode für Photonen- und Elektronenstrahlung – Teil 2: Dosimetrie hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern. 2020;
-
- Earley L, Moeller J, O’Rear J, Leavitt DD (1995) A method for total skin electron treatment for infants. Med Dosim 20(4):243–248 - DOI
-
- Elsayad K, Kroeger K, Greve B et al (2020) Low-dose total skin electron beam therapy: Quality of life improvement and clinical impact of maintenance and adjuvant treatment in patients with mycosis fungoides or Sezary syndrome. Strahlenther Onkol 196(1):77–84 - DOI
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials