Leadless pacemaker implantation: An unexpected complication
- PMID: 34743338
- DOI: 10.1111/pace.14400
Leadless pacemaker implantation: An unexpected complication
Abstract
The leadless pacemaker is an emerging technology with high efficacy and reduced complications rates. However, due to its novel status, some pitfalls remain to be addressed. We report the case of a 91-year-old patient undergoing a Micra pacemaker implantation. During the procedure, the maneuvers required for the adequate deployment of the device led to damaging of the septal tricuspid leaflet, resulting in severe tricuspid regurgitation. This is a severe mechanical complication of the Micra implantation technique, not previously reported in literature. In light of the novelty of the leadless pacemaker, we should remain cautious with regards to potential unreported complications.
Keywords: instrumentation; new technology; pacing.
© 2021 Wiley Periodicals LLC.
References
REFERENCES
-
- Roberts PR, Clementy N, Al Samadi F, et al. A leadless pacemaker in the real-world setting: the Micra transcatheter pacing system post-approval registry. Heart Rhythm. 2017; 14: 1375-1379.
-
- Reynolds D, Duray GZ, Omar R, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016; 374: 533-541.
-
- Ngo L, Nour D, Denman RA, et al. Safety and efficacy of leadless pacemakers: a systematic review and meta-analysis. J Am Heart Assoc. 2021; 10: e019212.
-
- Pagan E, Gabriels J, Khodak A, et al. Safety of leadless pacemaker implantation in the very elderly. Heart Rhythm. 2020; 17: 2023-2028.
-
- Tachibana M, Banba K, Matsumoto K, Ohara M. The feasibility of leadless pacemaker implantation for superelderly patients. Pacing Clin Electrophysiol. 2020; 43: 374-381.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical