. 2022 Feb 1;46(1):54-60.
doi: 10.1097/PXR.0000000000000053.

Effects of a modified passive socket system on short-term changes in residuum volume and comfort: A preliminary study in transtibial amputees

Affiliations

Effects of a modified passive socket system on short-term changes in residuum volume and comfort: A preliminary study in transtibial amputees

Huthaifa Atallah et al. Prosthet Orthot Int. .

Abstract

Background: Changes in residuum volume are a common problem in lower-limb amputees during prosthesis usage, and can lead to poor suspension, impaired gait, and tissue damage. Residuum volume can be affected by the in-socket air pressure, which will influence fluid flow in and around the residuum. The use of "active" pumps to reduce air pressure has been shown to conserve the residuum volume, but these are expensive and unlikely to be widely available. An alternative, passive approach, based on Boyles' law, is to introduce a larger distal void volume at the end of the socket and hence reduce the change in pressure for a given change in volume.

Objectives: To compare the performance across three test conditions (passive-conventional, with standard distal void; passive-with increased distal void; and active system) in terms of residuum volume changes and comfort.

Study design: Repeated-measures experiment under three test conditions.

Methods: Five transtibial amputee participants (three males and two females), aged between 27 and 67 years, and of mobility grade K2 or K3, were fitted with a bespoke test prosthesis that was adapted to include the three test conditions. Residuum volume was measured before and after walking under each test condition (presented in a random order). Comfort was also assessed after walking with each test condition.

Results: The reduction in residuum volume, relative to the baseline volume, was higher for the conventional passive system (4.2% ± 2.8%) compared with the modified passive (1.4% ± 1.4%) and active (1.6% ± 1.1%) systems.

Conclusion: The use of a passive suspension system with an increased distal void within the socket may help to stabilize the residuum volume during prosthesis usage.

PubMed Disclaimer

References

    1. Sanders JE, Cagle JC, Allyn KJ, et al. How do walking, standing, and resting influence transtibial amputee residual limb fluid volume? J Rehabil Res Dev 2014; 51: 201–212.
    1. Beil TL, Street GM, Covey SJ. Interface pressures during ambulation using suction and vacuum-assisted prosthetic sockets. J Rehabil Res Dev 2002; 39: 693–700.
    1. Sanders JE, Harrison DS, Myers TR, et al. Effects of elevated vacuum on in-socket residual limb fluid volume: case study results using bioimpedance analysis. J Rehabil Res Dev 2011; 48: 1231–1248.
    1. Eshraghi A, Abu Osman NA, Karimi M, et al. Gait biomechanics of individuals with transtibial amputation: effect of suspension system. PLoS One 2014; 9: e96988.
    1. Samitier CB, Guirao L, Costea M, et al. The benefits of using a vacuum-assisted socket system to improve balance and gait in elderly transtibial amputees. Prosthet Orthot Int 2016; 40: 83–88.

LinkOut - more resources