Test-Retest Reliability of a New Device Versus a Long-Arm Goniometer to Evaluate Knee Proprioception
- PMID: 34784583
- DOI: 10.1123/jsr.2021-0146
Test-Retest Reliability of a New Device Versus a Long-Arm Goniometer to Evaluate Knee Proprioception
Abstract
Context: Many methods used to evaluate knee proprioception have shortcomings that limit their use in clinical settings. Based on an inexpensive 3D camera, a new portable device was recently used to evaluate the joint position sense (JPS) of the knee joint. However, the test-retest reliability of the new method remains unclear. This study aimed to evaluate the test-retest reliability of the new device and a long-arm goniometer for assessing knee JPS, and to compare the variability of the 2 methods.
Design: Prospective observational study of the test-retest reliability of knee JPS measurements.
Methods: Twenty-one healthy adults were tested in 2 sessions with a 1-week interval. Three target knee flexion angles (30°, 45°, and 60°) were reproduced in each session. Target and reproduced angles were measured with both methods. Intraclass correlation coefficients, standard error of the measurement, and Bland-Altman plots were used to quantify test-retest reliability. Paired t tests were used to compare knee JPS (absolute error of the target-reproduced angle) between the methods.
Results: The new device (good to excellent intraclass correlation coefficients .74-.80; standard error of the measurement 0.52°-0.61°) demonstrated better test-retest reliability than the goniometer (poor to fair intraclass correlation coefficients .23-.43; standard error of the measurement 0.89°-2.07°) and better test-retest agreement (respective mean differences for the 30°, 45°, and 60° knee angles: 0.11°, 0.13°, and 0.41° for the new system; 0.84°, 1.52°, and 1.18° for the goniometer). The measurements (absolute errors of the target-reproduced angles) with the goniometer were significantly greater than those with the new device (P < .05); the SDs of repeated measurements with the goniometer (1.50°-2.41°) were greater than with the new device (1.08°-1.38°).
Conclusions: Given that the new device has good reliability and sufficient precision, it is the better alternative for evaluating knee JPS. Goniometers should be used with caution to assess knee JPS.
Keywords: agreement; joint position sense; method; motion capture system; reproducibility.
Similar articles
-
Test-retest reliability of joint position sense in the elbow among healthy adults.J Hand Ther. 2024 Apr-Jun;37(2):243-249. doi: 10.1016/j.jht.2023.08.015. Epub 2023 Sep 30. J Hand Ther. 2024. PMID: 37783630
-
Test-Retest Reliability of a Passive Joint Position Sense Test After ACL Reconstruction: Influence of Direction, Target Angle, Limb, and Outcome Measures.Orthop J Sports Med. 2023 Mar 21;11(3):23259671231157351. doi: 10.1177/23259671231157351. eCollection 2023 Mar. Orthop J Sports Med. 2023. PMID: 36970320 Free PMC article.
-
Interexaminer, intraexaminer and test-retest reliability of clinical knee joint-position-sense measurements using an image-capture technique.J Sport Rehabil. 2015 May 20;24(2):2013-0134. doi: 10.1123/jsr.2013-0134. Print 2015 May 1. J Sport Rehabil. 2015. PMID: 25310683
-
Intratester and intertester reliability and criterion validity of the parallelogram and universal goniometers for active knee flexion in healthy subjects.Physiother Res Int. 1997;2(3):150-66. doi: 10.1002/pri.97. Physiother Res Int. 1997. PMID: 9421820 Review.
-
Comparison of proprioception between osteoarthritic and age-matched unaffected knees: a systematic review and meta-analysis.Arch Orthop Trauma Surg. 2021 Mar;141(3):355-365. doi: 10.1007/s00402-020-03418-2. Epub 2020 Mar 30. Arch Orthop Trauma Surg. 2021. PMID: 32232620
Cited by
-
The effect of foot posture on static balance, ankle and knee proprioception in 18-to-25-year-old female student: a cross-sectional study.BMC Musculoskelet Disord. 2023 Jul 4;24(1):547. doi: 10.1186/s12891-023-06678-2. BMC Musculoskelet Disord. 2023. PMID: 37403076 Free PMC article.
-
A Reliability of Active and Passive Knee Joint Position Sense Assessment Using the Luna EMG Rehabilitation Robot.Int J Environ Res Public Health. 2022 Nov 29;19(23):15885. doi: 10.3390/ijerph192315885. Int J Environ Res Public Health. 2022. PMID: 36497961 Free PMC article.
-
Monitoring Cortical and Neuromuscular Activity: Six-month Insights into Knee Joint Position Sense Following ACL Reconstruction.Int J Sports Phys Ther. 2024 Nov 2;19(11):1290-1303. doi: 10.26603/001c.124840. eCollection 2024. Int J Sports Phys Ther. 2024. PMID: 39502546 Free PMC article.
-
Reliability of Sensorimotor Control Tests in Individuals with Adolescent Idiopathic Scoliosis.Muscles. 2024 Nov 15;3(4):376-392. doi: 10.3390/muscles3040032. Muscles. 2024. PMID: 40757521 Free PMC article.
-
Reliability of the active knee joint position sense test and influence of limb dominance and sex.Sci Rep. 2023 Jan 4;13(1):152. doi: 10.1038/s41598-022-26932-2. Sci Rep. 2023. PMID: 36599861 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical