Comparison of VividTrac, King Vision and Macintosh laryngoscopes in normal and difficult airways during simulated cardiopulmonary resuscitation among novices
- PMID: 34793558
- PMCID: PMC8601572
- DOI: 10.1371/journal.pone.0260140
Comparison of VividTrac, King Vision and Macintosh laryngoscopes in normal and difficult airways during simulated cardiopulmonary resuscitation among novices
Abstract
Introduction: Early endotracheal intubation improves neurological outcomes in cardiopulmonary resuscitation, although cardiopulmonary resuscitation is initially carried out by personnel with limited experience in a significant proportion of cases. Videolaryngoscopes might decrease the number of attempts and time needed, especially among novices. We sought to compare videolaryngoscopes with direct laryngoscopes in simulated cardiopulmonary resuscitation scenarios.
Materials and methods: Forty-four medical students were recruited to serve as novice users. Following brief, standardized training, students executed endotracheal intubation with the King Vision®, Macintosh and VividTrac® laryngoscopes, on a cardiopulmonary resuscitation trainer in normal and difficult airway scenarios. We evaluated the time to and proportion of successful intubation, the best view of the glottis, esophageal intubation, dental trauma and user satisfaction.
Results: In the normal airway scenario, significantly shorter intubation times were achieved using the King Vision® than the Macintosh laryngoscope. In the difficult airway scenario, we found that the VividTrac® was superior to the King Vision® and Macintosh laryngoscopes in the laryngoscopy time. In both scenarios, we noted no difference in the first-attempt success rate, but the best view of the glottis and dental trauma, esophageal intubation and bougie use were more frequent with the Macintosh laryngoscope than with the videolaryngoscopes. The shortest tube insertion times were achieved using the King Vision® in both scenarios.
Conclusion: All providers achieved successful intubation within three attempts, but we found no device superior in any of our scenarios regarding the first-attempt success rate. The King Vision® was superior to the Macintosh laryngoscope in the intubation time in the normal airway scenario and noninferior in the difficult airway scenario for novice users. We noted significantly less esophageal intubation using the videolaryngoscopes than using the Macintosh laryngoscope in both scenarios. Based on our results, the KingVision® might be recommended over the VividTrac® and Macintosh laryngoscopes for further evaluation.
Conflict of interest statement
The authors have declared that no competing interests exist.
Similar articles
-
Comparison of VividTrac®, Airtraq®, King Vision®, Macintosh Laryngoscope and a Custom-Made Videolaryngoscope for difficult and normal airways in mannequins by novices.BMC Anesthesiol. 2017 May 26;17(1):68. doi: 10.1186/s12871-017-0362-y. BMC Anesthesiol. 2017. PMID: 28549421 Free PMC article.
-
Comparison of the UEScope videolaryngoscope with the Macintosh laryngoscope during simulated cardiopulmonary resuscitation: A randomized, cross-over, multi-center manikin study.Medicine (Baltimore). 2018 Sep;97(36):e12085. doi: 10.1097/MD.0000000000012085. Medicine (Baltimore). 2018. PMID: 30200089 Free PMC article. Clinical Trial.
-
Comparison of Macintosh Laryngoscope, King Vision®, VividTrac®, AirAngel Blade®, and a Custom-Made 3D-Printed Video Laryngoscope for Difficult and Normal Airways in Mannequins by Novices-A Non-Inferiority Trial.J Clin Med. 2024 May 30;13(11):3213. doi: 10.3390/jcm13113213. J Clin Med. 2024. PMID: 38892925 Free PMC article.
-
Videolaryngoscope designs for tracheal intubation in adults: a systematic review with network meta-analysis of randomised controlled trials.Anaesthesia. 2025 Jul;80(7):823-833. doi: 10.1111/anae.16597. Epub 2025 Mar 11. Anaesthesia. 2025. PMID: 40069118
-
Videolaryngoscope versus Macintosh laryngoscope for tracheal intubation in adults with obesity: A systematic review and meta-analysis.J Clin Anesth. 2018 Feb;44:69-75. doi: 10.1016/j.jclinane.2017.11.008. Epub 2017 Nov 20. J Clin Anesth. 2018. PMID: 29156438
Cited by
-
Comparative study of the McGrath™ videolaryngoscope blades and conventional laryngoscopy efficacy during mechanical chest compressions: Insights from a randomized trial with 90 anesthesiologists on objective and subjective parameters.PLoS One. 2024 Sep 20;19(9):e0310796. doi: 10.1371/journal.pone.0310796. eCollection 2024. PLoS One. 2024. PMID: 39302951 Free PMC article. Clinical Trial.
-
Airway management during ongoing chest compressions-direct vs. video laryngoscopy. A randomised manikin study.PLoS One. 2023 Feb 9;18(2):e0281186. doi: 10.1371/journal.pone.0281186. eCollection 2023. PLoS One. 2023. PMID: 36757942 Free PMC article. Clinical Trial.
References
-
- van Schuppen H, Boomars R, Kooij FO, den Tex P, Koster RW, Hollmann MW. Optimizing airway management and ventilation during prehospital advanced life support in out-of-hospital cardiac arrest: A narrative review. Best Pract Res Clin Anaesthesiol. 2021;35: 67–82. doi: 10.1016/j.bpa.2020.11.003 - DOI - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical