. 2022 Apr;87(4):1952-1970.
doi: 10.1002/mrm.29094. Epub 2021 Nov 23.

B1 inhomogeneity correction of RARE MRI at low SNR: Quantitative in vivo 19 F MRI of mouse neuroinflammation with a cryogenically-cooled transceive surface radiofrequency probe

Affiliations

B1 inhomogeneity correction of RARE MRI at low SNR: Quantitative in vivo 19 F MRI of mouse neuroinflammation with a cryogenically-cooled transceive surface radiofrequency probe

Paula Ramos Delgado et al. Magn Reson Med. 2022 Apr.

Abstract

Purpose: Low SNR in fluorine-19 (19 F) MRI benefits from cryogenically-cooled transceive surface RF probes (CRPs), but strong B1 inhomogeneities hinder quantification. Rapid acquisition with refocused echoes (RARE) is an SNR-efficient method for MRI of neuroinflammation with perfluorinated compounds but lacks an analytical signal intensity equation to retrospectively correct B1 inhomogeneity. Here, a workflow was proposed and validated to correct and quantify 19 F-MR signals from the inflamed mouse brain using a 19 F-CRP.

Methods: In vivo 19 F-MR images were acquired in a neuroinflammation mouse model with a quadrature 19 F-CRP using an imaging setup including 3D-printed components to acquire co-localized anatomical and 19 F images. Model-based corrections were validated on a uniform 19 F phantom and in the neuroinflammatory model. Corrected 19 F-MR images were benchmarked against reference images and overlaid on in vivo 1 H-MR images. Computed concentration uncertainty maps using Monte Carlo simulations served as a measure of performance of the B1 corrections.

Results: Our study reports on the first quantitative in vivo 19 F-MR images of an inflamed mouse brain using a 19 F-CRP, including in vivo T1 calculations for 19 F-nanoparticles during pathology and B1 corrections for 19 F-signal quantification. Model-based corrections markedly improved 19 F-signal quantification from errors > 50% to < 10% in a uniform phantom (p < 0.001). Concentration uncertainty maps ex vivo and in vivo yielded uncertainties that were generally < 25%. Monte Carlo simulations prescribed SNR ≥ 10.1 to reduce uncertainties < 10%, and SNR ≥ 4.25 to achieve uncertainties < 25%.

Conclusion: Our model-based correction method facilitated 19 F signal quantification in the inflamed mouse brain when using the SNR-boosting 19 F-CRP technology, paving the way for future low-SNR 19 F-MRI applications in vivo.

Keywords: 19F-MRI; B1 correction; RARE; inflammation; transceive surface RF probe.

PubMed Disclaimer

References

REFERENCES

    1. Ruiz-Cabello J, Barnett B, Bottomley P, Bulte J. Fluorine (19F) MRS and MRI in biomedicine. NMR Biomed. 2011;24:114-129.
    1. Waiczies S, Ji Y, Niendorf T. Tracking methods for dendritic cells. In: Flögel U, Ahrens E, eds. Fluorine Magnetic Resonance Imaging. Pan Stanford Publishing; 2017:243-281.
    1. Chen J, Lanza G, Wickline S. Quantitative magnetic resonance fluorine imaging: today and tomorrow. Wiley Interdiscip Rev Nanomed Nanobiotechnol. 2015;2:431-440.
    1. Axel L, Hayes C. Surface coil magnetic resonance imaging. Archives Internationales De Physiologie Et De Biochimie. 1985;93:11-18.
    1. Ratering D, Baltes C, Nordmeyer-Massner J, Marek D, Rudin M. Performance of a 200-MHz cryogenic RF probe designed for MRI and MRS of the murine brain. Magn Reson Med. 2008;59:1440-1447.

Publication types

LinkOut - more resources