Cochlear Implant Compression Optimization for Musical Sound Quality in MED-EL Users
- PMID: 34812791
- DOI: 10.1097/AUD.0000000000001145
Cochlear Implant Compression Optimization for Musical Sound Quality in MED-EL Users
Abstract
Objectives: Variations in loudness are a fundamental component of the music listening experience. Cochlear implant (CI) processing, including amplitude compression, and a degraded auditory system may further degrade these loudness cues and decrease the enjoyment of music listening. This study aimed to identify optimal CI sound processor compression settings to improve music sound quality for CI users.
Design: Fourteen adult MED-EL CI recipients participated (Experiment No. 1: n = 17 ears; Experiment No. 2: n = 11 ears) in the study. A software application using a modified comparison category rating (CCR) test method allowed participants to compare and rate the sound quality of various CI compression settings while listening to 25 real-world music clips. The two compression settings studied were (1) Maplaw, which informs audibility and compression of soft level sounds, and (2) automatic gain control (AGC), which applies compression to loud sounds. For each experiment, one compression setting (Maplaw or AGC) was held at the default, while the other was varied according to the values available in the clinical CI programming software. Experiment No. 1 compared Maplaw settings of 500, 1000 (default), and 2000. Experiment No. 2 compared AGC settings of 2.5:1, 3:1 (default), and 3.5:1.
Results: In Experiment No. 1, the group preferred a higher Maplaw setting of 2000 over the default Maplaw setting of 1000 (p = 0.003) for music listening. There was no significant difference in music sound quality between the Maplaw setting of 500 and the default setting (p = 0.278). In Experiment No. 2, a main effect of AGC setting was found; however, no significant difference in sound quality ratings for pairwise comparisons were found between the experimental settings and the default setting (2.5:1 versus 3:1 at p = 0.546; 3.5:1 versus 3:1 at p = 0.059).
Conclusions: CI users reported improvements in music sound quality with higher than default Maplaw or AGC settings. Thus, participants preferred slightly higher compression for music listening, with results having clinical implications for improving music perception in CI users.
Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
The authors have no conflicts of interest to disclose.
References
-
- Bates D., Maechler M., Bolker B., Walker S. (2015). Fitting linear mixed-effects models using lme4. J Stat Softw, 67, 1–48.
-
- Boyd P. J. (2006). Effects of programming threshold and maplaw settings on acoustic thresholds and speech discrimination with the MED-EL COMBI 40+ cochlear implant. Ear Hear, 27, 608–618.
-
- Boyd P., Euthymiades A. (2009). Comparison of loudness adjustments by MCL and maplaw in users of the MED-EL COMBI 40/40+ cochlear implant system. Cochlear Implants Int, 10, 203–217.
-
- Buyens W., Moonen M., Wouters J., van Dijk B. (2017). A model for music complexity applied to music preprocessing for cochlear implants. 25th European Signal Processing Conference (EUSIPCO). 971–975. doi: 10.23919/EUSIPCO.2017.8081352
-
- Buyens W., van Dijk B., Moonen M., Wouters J. (2014). Music mixing preferences of cochlear implant recipients: A pilot study. Int J Audiol, 53, 294–301.
REFERENCE NOTE
-
- Katelyn Glassman & Josh Stohl, MED-EL Corp., personal communication, November 20, 2017
MeSH terms
LinkOut - more resources
Full Text Sources
Medical