Outcomes of hybrid closed-loop insulin delivery activated 24/7 versus evening and night in free-living prepubertal children with type 1 diabetes: A multicentre, randomized clinical trial
- PMID: 34816597
- DOI: 10.1111/dom.14605
Outcomes of hybrid closed-loop insulin delivery activated 24/7 versus evening and night in free-living prepubertal children with type 1 diabetes: A multicentre, randomized clinical trial
Abstract
Aim: To assess the safety and efficacy of hybrid closed-loop (HCL) insulin delivery 24/7 versus only evening and night (E/N), and on extended 24/7 use, in free-living children with type 1 diabetes.
Materials and methods: Prepubertal children (n = 122; 49 females/73 males; age, 8.6 ± 1.6 years; diabetes duration, 5.2 ± 2.3 years; insulin pump use, 4.6 ± 2.5 years; HbA1c 7.7% ± 0.7%/61 ± 5 mmol/mol) from four centres were randomized for 24/7 versus E/N activation of the Tandem Control-IQ system for 18 weeks. Afterwards, all children used the activated system 24/7 for 18 more weeks. The primary outcome was the percentage of time spent in the 70-180 mg/dL glucose range (TIR).
Results: HCL was active 94.1% and 51.1% of the time in the 24/7 and E/N modes, respectively. TIR from baseline increased more in the 24/7 versus the E/N mode (52.9% ± 9.5% to 67.3% ± 5.6% [+14.4%, 95% CI 12.4%-16.7%] vs. 55.1% ± 10.8% to 64.7% ± 7.0% [+9.6%, 95% CI 7.4%-11.6%]; P = .001). Mean percentage time below range was similarly reduced, from 4.2% and 4.6% to 2.7%, and the mean percentage time above range decreased more in the 24/7 mode (41.9% to 30.0% [-11.9%, 95% CI 9.7%-14.6%] vs. 39.8% to 32.6% [-7.2%, 95% CI 5.0%-9.9%]; P = .007). TIR increased through the whole range of baseline levels and always more with 24/7 use. The results were maintained during the extension phase in those initially on 24/7 use and improved in those with initial E/N use up to those with 24/7 use. Neither ketoacidosis nor severe hypoglycaemia occurred.
Conclusions: The current study shows the safety and efficacy of the Tandem Control-IQ system in free-living children with type 1 diabetes for both E/N and 24/7 use; 24/7 use shows better outcomes, sustained for up to 36 weeks with no safety issues.
Keywords: children; clinical trial; closed-loop; glucose control; type 1 diabetes.
© 2021 John Wiley & Sons Ltd.
References
REFERENCES
-
- Foster NC, Beck RW, Miller KM, et al. State of type 1 diabetes management and outcomes from the T1D exchange in 2016-2018. Diabetes Technol Ther. 2019;21(2):66-72.
-
- Karageorgiou V, Papaioannou TG, Bellos I, et al. Effectiveness of artificial pancreas in the non-adult population: a systematic review and network meta-analysis. Metabolism. 2019;90:20-30.
-
- Thabit H, Hovorka R. Coming of age: the artificial pancreas for type 1 diabetes. Diabetologia. 2016;59(9):1795-1805.
-
- Lal RA, Ekhlaspour L, Hood K, Buckingham B. Realizing a closed-loop (artificial pancreas) system for the treatment of type 1 diabetes. Endocr Rev. 2019;40(6):1521-1546.
-
- Boughton CK, Hartnell S, Allen JM, Hovorka R. The importance of prandial insulin bolus timing with hybrid closed-loop systems. Diabet Med. 2019;36(12):1716-1717.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Research Materials