Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 2022 Apr;17(4):795-804.
doi: 10.1007/s11548-021-02520-z. Epub 2021 Nov 24.

Telemedical percussion: objectifying a fundamental clinical examination technique for telemedicine

Affiliations
Review

Telemedical percussion: objectifying a fundamental clinical examination technique for telemedicine

Roman Krumpholz et al. Int J Comput Assist Radiol Surg. 2022 Apr.

Abstract

Purpose: While demand for telemedicine is increasing, patients are currently restricted to tele-consultation for the most part. Fundamental diagnostics like the percussion still require the in person expertize of a physician. To meet today's challenges, a transformation of the manual percussion into a standardized, digital version, ready for telemedical execution is required.

Methods: In conjunction with a comprehensive telemedical diagnostic system, in which patients can get examined by a remote-physician, a series of three robotic end-effectors for mechanical percussion were developed. Comprising a motor, a magnetic and a pneumatic-based version, the devices strike a pleximeter to perform the percussion. Emitted sounds were captured using a microphone-equipped stethoscope. The 84 recordings were further integrated into a survey in order to classify lung and non-lung samples.

Results: The study with 21 participants comprised physicians, medical students and non-medical-related raters in equal parts. With 71.4% correctly classified samples, the ventral motorized device prevailed. While the result is significantly better compared to a manual or pneumatic percussion in this very setup, it only has a small edge over the magnetic devices. In addition, for all ventral versions non-lung regions were rather correctly identified than lung regions.

Conclusion: The overall setup proves the feasibility of a telemedical percussion. Despite the fact, that produced sounds differ compared to today's manual technique, the study shows that a standardized mechanical percussion has the potential to improve the gold standard's accuracy. While further extensive medical evaluation is yet to come, the system paves the way for future uncompromised remote examinations.

Keywords: Auscultation; Percussion; Robotic examination; Robotic percussion; Tele-diagnostic; Telemedicine.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no conflict of interest.

Figures

Fig. 1
Fig. 1
Left: Ivory pelximeter with handles used to transfer the strikes during a percussion onto the patient’s tissue [8]. Right: Sibson’s percussor-pleximeter, a percussion device with an integrated pleximeter, allowing for higher standardized diagnostics [8]
Fig. 2
Fig. 2
Left: Photograph of the examination cabin. White circles indicate the distinct examination modules. Right: Detail shot, showing the revolver-like fixture at the robotic arm of module 3 with the attached palpation and auscultation device, as well as the magnetic percussion end-effector during evaluation
Fig. 3
Fig. 3
Inner workings of the three different percussion instruments. Picture a shows the spring-loaded motorized version, picture b depicts the transistor controlled solenoid and picture c displays the pneumatic percussion instrument, driven by an external supply of compressed air
Fig. 4
Fig. 4
Percussion points on the ventral (left) and dorsal (right) patient side. The lung region is highlighted. Auscultation was always carried out dosal and in close proximity to the percussion end-effector
Fig. 5
Fig. 5
Device abbreviations—MAN manual percussion, MOT-D mechanic device with dorsal percussion and auscultation, MOT-V mechanic device with ventral percussion and dorsal auscultation, MAGN-D magnetic device with dorsal percussion and auscultation, MAGN-V magnetic device with ventral percussion and dorsal auscultation, PNEU-D pneumatic device with dorsal percussion and auscultation, PNEU-V pneumatic device with ventral percussion and dorsal auscultation, the color scheme from A and B is also used in C, in B the symbols mark the mean and the whiskers the standard deviation on the respective axis/**p value < 0.01, *p value < 0.05, n.s. p value 0.05 in Friedman test with Wilcoxon rank sum test as post-hoc analysis, p values are adjusted by Bonferroni correction

References

    1. Auenbrugger L (1761) Inventum novum, ex percus-sione thoracis humani ut signo abstrusos interni pectoris morbos detegendi, vol 1. J.T.Trattner, Vienne, p 525
    1. Ayodele K, Ogunlade O, Olugbon O, Akinwale O, Kehinde L. A medical percussion instrument using a wavelet-based method for archivable output and automatic classification. Comput Biol Med. 2020;127:104100. doi: 10.1016/j.compbiomed.2020.104100. - DOI - PubMed
    1. Bohadana A, Kraman S. Consistency of sternal percussion performed manually and with mechanical thumper. Eur Respir J. 1992;5(8):1004–1008. - PubMed
    1. Bohadana AB, Kraman SS. Transmission of sound generated by sternal percussion. J Appl Physiol. 1989;66(1):273–277. doi: 10.1152/jappl.1989.66.1.273. - DOI - PubMed
    1. Bohadana AB, Patel R, Kraman SS. Contour maps of auscultatory percussion in healthy subjects and patients with large intrapulmonary lesions. Lung. 1989;167(1):359–372. doi: 10.1007/BF02714964. - DOI - PubMed