Successful implantation of S-ICD using the intermuscular two-incision technique in a patient with severe pectus excavatum
- PMID: 34853691
- PMCID: PMC8617332
- DOI: 10.1002/ccr3.5143
Successful implantation of S-ICD using the intermuscular two-incision technique in a patient with severe pectus excavatum
Abstract
A patient with severe pectus excavatum, dilated ischemic heart disease, and indication for defibrillator implantation for primary prevention of sudden death underwent successful ultrasound-guided serratus anterior plane block and parasternal block with intermuscular two-incision technique implantation with no complications. At follow-up, all the parameters resulted stable with excellent signal sensing.
Keywords: pectus excavatum; serratus anterior plane block; subcutaneous implantable cardioverter defibrillator; two‐incision technique.
© 2021 The Authors. Clinical Case Reports published by John Wiley & Sons Ltd.
Conflict of interest statement
Dr. Andrea Droghetti is a consultant for Boston Scientific. The other authors declare that they have no conflicts of interest.
Figures



References
-
- Burke MC, Gold MR, Knight BP, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2‐year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol. 2015;65(16):1605‐1615. - PubMed
-
- Rordorf R, Casula M, Pezza L, et al. Subcutaneous versus transvenous implantable defibrillator: an updated meta‐analysis. Fortuni F, Sanzo A, Savastano S, Vicentini A. Heart Rhythm. 2020;16:382‐391. - PubMed
-
- Droghetti A, Basso Ricci E, Scimia P, Harizai F, Marini M. Ultrasound‐guided serratus anterior plane block combined with the two‐incision technique for subcutaneous ICD implantation. Pacing Clin Electrophysiol. 2018;41:517‐523. - PubMed
-
- Droghetti A, Fusco P, Marini M, et al. Ultrasound‐guided serratus anterior plane block and parasternal block in cooperative sedation for S‐ICD implantation. Pacing Clin Electrophysiol. 2019;42(7):1076‐1078. - PubMed
-
- Ziacchi M, Bisignani G, Palmisano P, et al. Serratus anterior plane block in subcutaneous implantable cardioverter defibrillator implantation: a case‐control analysis. J Cardiovasc Electrophysiol. 2020;31(1):144‐149. - PubMed