Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2021 Dec 2;5(1):51.
doi: 10.1186/s41747-021-00251-z.

Detecting low blood concentrations in joints using T1 and T2 mapping at 1.5, 3, and 7 T: an in vitro study

Affiliations

Detecting low blood concentrations in joints using T1 and T2 mapping at 1.5, 3, and 7 T: an in vitro study

Flora H P van Leeuwen et al. Eur Radiol Exp. .

Abstract

Background: Intra-articular blood causes irreversible joint damage, whilst clinical differentiation between haemorrhagic joint effusion and other effusions can be challenging. An accurate non-invasive method for the detection of joint bleeds is lacking. The aims of this phantom study were to investigate whether magnetic resonance imaging (MRI) T1 and T2 mapping allows for differentiation between simple and haemorrhagic joint effusion and to determine the lowest blood concentration that can be detected.

Methods: Solutions of synovial fluid with blood concentrations ranging from 0 to 100% were scanned at 1.5, 3, and 7 T. T1 maps were generated with an inversion recovery technique and T2 maps from multi spin-echo sequences. In both cases, the scan acquisition times were below 5 min. Regions of interest were manually drawn by two observers in the obtained T1 and T2 maps for each sample. The lowest detectable blood concentration was determined for all field strengths.

Results: At all field strengths, T1 and T2 relaxation times decreased with higher blood concentrations. The lowest detectable blood concentrations using T1 mapping were 10% at 1.5 T, 25% at 3 T, and 50% at 7 T. For T2 mapping, the detection limits were 50%, 5%, and 25%, respectively.

Conclusions: T1 and T2 mapping can detect different blood concentrations in synovial fluid in vitro at clinical field strengths. Especially, T2 measurements at 3 T showed to be highly sensitive. Short acquisition times would make these methods suitable for clinical use and therefore might be promising tools for accurate discrimination between simple and haemorrhagic joint effusion in vivo.

Keywords: Haemarthrosis; Image interpretation (computer-assisted); Magnetic resonance imaging; Phantoms (imaging); Synovial fluid.

PubMed Disclaimer

Conflict of interest statement

The authors of this manuscript declare relationships with the following companies: WF has received research grants from NovoNordisk and Pfizer which were paid to the institution. LV received a research grant from CSL Behring which was paid to the institution and reports to be in the advisory boards of Swedish Orphan Biovitrum BV (sobi), Tremeau Pharmaceuticals and CSL Behring. KF has received speaker's fees from Bayer, Baxter/Shire, Biotest, CSL Behring, Octapharma, Pfizer, NovoNordisk; performed consultancy for Baxter/Shire, Biogen, CSL-Behring, Freeline, NovoNordisk, Pfizer, Roche and SOBI; and has received research support from Bayer, Pfizer, Baxter/Shire, and Novo Nordisk. PJ reported to perform consultancy for Sanifit and Inozyme. The Radiology department of the University Medical Center Utrecht received research support from Philips.

Figures

Fig. 1
Fig. 1
a Nuclear magnetic resonance tubes with different ratios of synovial fluid and blood. Blood percentage from left to right: 0%, 2.5%, 5%, 10%, 25%, 50%, 75%, and 100%. b Phantom setup: a cylindrical water-filled holder with six samples, horizontally placed in a 16-channel knee coil using a 3-T MRI system. The phantom was equipped with a fiberoptic thermometer (the asterisk in all images) and a heating system based on heat exchange between water in the heating system (black arrows) and the water or oil volume in the phantom. c Scan image from the T1 mapping sequence. For each tube, a region of interest was manually placed within the tube region (example in yellow). d Scan image from the T2 mapping sequence
Fig. 2
Fig. 2
Mean T1 and T2 value estimates for the different blood concentrations, at 3 field strengths. For 7 T, some of the estimates were not available, because artefacts or relaxation times were too short to be measured
Fig. 3
Fig. 3
Effect of different acceleration techniques on T2 estimates at 1.5 T. T2 exhibited an inverse dependence on the blood concentration with all acceleration techniques. The three accelerated T2 mapping methods showed good agreement with the reference scan (echo spacing 30 ms), except from the 5% blood concentration measurement using the EPI acceleration technique. EPI, echo-planar imaging; TSE, turbo spin-echo; TSE+SENSE, turbo spin-echo with sensitivity encoding
Fig. 4
Fig. 4
Bland-Altman plots for interobserver agreement regarding T1 and T2 measurements at 1.5 and 3 T. Horizontal dashed lines indicate the upper and lower limit of agreement from the mean by the two observers (ULoA/LLoA). For T1 measurements, the mean was − 28.99 ms, with a limit of agreement (LoA) equal to 143 ms. For T2 measurements, the mean was − 1.164 ms with LoA equal to 148 ms

References

    1. Pulles AE, Mastbergen SC, Schutgens REG, et al. Pathophysiology of hemophilic arthropathy and potential targets for therapy. Pharmacol Res. 2017;115:192–199. doi: 10.1016/j.phrs.2016.11.032. - DOI - PubMed
    1. Jansen NWD, Roosendaal G, Lafeber FPJG. Understanding haemophilic arthropathy: an exploration of current open issues. Br J Haematol. 2008;143:632–640. doi: 10.1111/j.1365-2141.2008.07386.x. - DOI - PubMed
    1. Jansen NWD, Roosendaal G, Bijlsma JWJ, et al. Exposure of human cartilage tissue to low concentrations of blood for a short period of time leads to prolonged cartilage damage: an in vitro study. Arthritis Rheum. 2007;56:199–207. doi: 10.1002/art.22304. - DOI - PubMed
    1. Timmer MA, Pisters MF, de Kleijn P, et al. Differentiating between signs of intra-articular joint bleeding and chronic arthropathy in haemophilia: a narrative review of the literature. Haemophilia. 2015;21:289–296. doi: 10.1111/hae.12667. - DOI - PubMed
    1. Ceponis A, Wong-Sefidan I, Glass CS, von Drygalski A. Rapid musculoskeletal ultrasound for painful episodes in adult haemophilia patients. Haemophilia. 2013;19:790–798. doi: 10.1111/hae.12175. - DOI - PubMed

Publication types

MeSH terms

LinkOut - more resources