. 2022 Jan:146:105046.
doi: 10.1016/j.jcv.2021.105046. Epub 2021 Nov 25.

Evaluating sensitivity and specificity of the Biomeme Franklin™ three9 real-time PCR device and SARS-CoV-2 go-strips assay using clinical samples

Affiliations

Evaluating sensitivity and specificity of the Biomeme Franklin™ three9 real-time PCR device and SARS-CoV-2 go-strips assay using clinical samples

Chet R Voelker et al. J Clin Virol. 2022 Jan.

Abstract

We evaluated the sensitivity and specificity of the Biomeme Franklin™ three9 Real-Time PCR Thermocycler and Biomeme SARS-CoV-2 Go-Strips in the detection of SARS-CoV-2. The Biomeme Franklin™ three9 platform is a portable, battery-operated system that could be used in remote settings. We assessed performance of the Biomeme SARS-CoV-2 detection system at a wide range of viral concentrations, examined cross-reactivity of the SARS-CoV-2 Go-Strips against several near-neighbor respiratory pathogens, and evaluated agreement against the BioFire® Respiratory Panel 2.1 in four clinical sample types. Our data indicate the Biomeme Go-Strips can reliably detect SARS-CoV-2 at a concentration of 4.2 × 103 copies/mL. No cross reactivity of the Go-Strips targets was detected against any of the tested near-neighbor respiratory pathogens. Cohen's kappa statistics ranged from 0.68 to 0.92 between results from the Biomeme SARS-CoV-2 Go-Strips and the BioFire® Respiratory Panel 2.1 in all the different sample types. Compared to the BioFire® Respiratory Panel 2.1, the Biomeme SARS-CoV-2 Go-Strips demonstrated statistically significantly lower sensitivity in 3 out of 5 sample types. Overall, our study demonstrates the Biomeme Franklin™ three9 used with the SARS-CoV-2 Go-Strips is an effective system for the detection of SARS-CoV-2 that could potentially be used in a remote or austere environment.

Keywords: Austere environment; COVID19; Diagnostic testing; Polymerase chain reaction; SARS-CoV-2.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

    1. WHO Director-General's opening Remarks At the Media Briefing On COVID-19 - 11 March 2020. 2020.
    1. WHO Coronavirus (COVID-19) Dashboard. 2021.
    1. Creager H.M., Cabrera B., Schnaubelt A., Cox J.L., Cushman-Vokoun A.M., Shakir S.M., et al. Clinical evaluation of the BioFire(R) Respiratory Panel 2.1 and detection of SARS-CoV-2. J. Clin. Virol. 2020;129 - PMC - PubMed
    1. Henning-Smith C. The Unique Impact of COVID-19 on Older Adults in Rural Areas. J. Aging Soc. Policy. 2020;32:396–402. - PubMed
    1. Michael N.L. SARS-CoV-2 in the U.S. military - lessons for civil society. N. Engl. J. Med. 2020;383:2472–2473. - PMC - PubMed

Publication types