Technical note: Does the greater power of pencil beam scanning reduce the need for a proton gantry? A study of head-and-neck and brain tumors
- PMID: 34919736
- DOI: 10.1002/mp.15409
Technical note: Does the greater power of pencil beam scanning reduce the need for a proton gantry? A study of head-and-neck and brain tumors
Abstract
Purpose: Proton therapy systems without a gantry can be more compact and less expensive in terms of capital cost and therefore more available to a larger patient population. Would the advances in pencil beam scanning (PBS) and robotics make gantry-less treatment possible? In this study, we explore if the high-quality treatment plans can be obtained without a gantry.
Methods and materials: We recently showed that proton treatments with the patient in an upright position may be feasible with a new soft robotic immobilization device and imaging which enables multiple possible patient orientations during a treatment. In this study, we evaluate if this new treatment geometry could enable high quality treatment plans without a gantry. We created PBS treatment plans for seven patients with head-and-neck or brain tumors. Each patient was planned with two scenarios: one with a gantry with the patient in supine position and the other with a gantry-less fixed horizontal beam-line with the patient sitting upright. For the treatment plans, dose-volume-histograms (DVHs), target homogeneity index (HI), mean dose, , and are reported. A robustness analysis of one plan was performed with 2.5-mm setup errors and 3.5% range uncertainties with nine scenarios.
Results: Most of the PBS-gantry-less plans had similar target HI and organs-at-risk mean dose as compared to PBS-gantry plans and similar robustness with respect to range uncertainties and setup errors.
Conclusions: PBS provides sufficient power to deliver high quality treatment plans without requiring a gantry for head-and-neck or brain tumors. In combination with the development of the new positioning and immobilization methods required to support this treatment geometry, this work suggests the feasibility of further development of a compact proton therapy system with a fixed horizontal beam-line to treat patients in sitting and reclined positions.
Keywords: compact proton therapy; fixed proton beam-line; pencil beam scanning (PBS); robotics.
© 2021 American Association of Physicists in Medicine.
References
REFERENCES
-
- Weber DC, Trofimov AV, Delaney TF, Bortfeld T. A treatment planning comparison of intensity modulated photon and proton therapy for paraspinal sarcomas. International Journal of Radiation Oncology* Biology* Physics. 2004;58:1596-1606.
-
- Lomax AJ, Bortfeld T, Goitein G, et al. A treatment planning inter-comparison of proton and intensity modulated photon radiotherapy. Radiother Oncol. 1999;51:257-271.
-
- Mock U, Georg D, Bogner J, Auberger T, Pötter R. Treatment planning comparison of conventional, 3D conformal, and intensity-modulated photon (IMRT) and proton therapy for paranasal sinus carcinoma. International Journal of Radiation Oncology* Biology* Physics. 2004;58:147-154.
-
- Zhang X, Li Y, Pan X, et al. Intensity-modulated proton therapy reduces the dose to normal tissue compared with intensity-modulated radiation therapy or passive scattering proton therapy and enables individualized radical radiotherapy for extensive stage IIIB non-small-cell lung cancer: a virtual clinical study. International Journal of Radiation Oncology* Biology* Physics. 2010;77:357-366.
-
- Safai S, Trofimov A, Adams J, Engelsman M, Bortfeld T. The rationale for intensity-modulated proton therapy in geometrically challenging cases. Phys Med Biol. 2013;58:6337.
MeSH terms