. 2022 Jan 18;24(1):e27487.
doi: 10.2196/27487.

Accuracy Assessment of Oura Ring Nocturnal Heart Rate and Heart Rate Variability in Comparison With Electrocardiography in Time and Frequency Domains: Comprehensive Analysis

Affiliations

Accuracy Assessment of Oura Ring Nocturnal Heart Rate and Heart Rate Variability in Comparison With Electrocardiography in Time and Frequency Domains: Comprehensive Analysis

Rui Cao et al. J Med Internet Res. .

Abstract

Background: Photoplethysmography is a noninvasive and low-cost method to remotely and continuously track vital signs. The Oura Ring is a compact photoplethysmography-based smart ring, which has recently drawn attention to remote health monitoring and wellness applications. The ring is used to acquire nocturnal heart rate (HR) and HR variability (HRV) parameters ubiquitously. However, these parameters are highly susceptible to motion artifacts and environmental noise. Therefore, a validity assessment of the parameters is required in everyday settings.

Objective: This study aims to evaluate the accuracy of HR and time domain and frequency domain HRV parameters collected by the Oura Ring against a medical grade chest electrocardiogram monitor.

Methods: We conducted overnight home-based monitoring using an Oura Ring and a Shimmer3 electrocardiogram device. The nocturnal HR and HRV parameters of 35 healthy individuals were collected and assessed. We evaluated the parameters within 2 tests, that is, values collected from 5-minute recordings (ie, short-term HRV analysis) and the average values per night sleep. A linear regression method, the Pearson correlation coefficient, and the Bland-Altman plot were used to compare the measurements of the 2 devices.

Results: Our findings showed low mean biases of the HR and HRV parameters collected by the Oura Ring in both the 5-minute and average-per-night tests. In the 5-minute test, the error variances of the parameters were different. The parameters provided by the Oura Ring dashboard (ie, HR and root mean square of successive differences [RMSSD]) showed relatively low error variance compared with the HRV parameters extracted from the normal interbeat interval signals. The Pearson correlation coefficient tests (P<.001) indicated that HR, RMSSD, average of normal heart beat intervals (AVNN), and percentage of successive normal beat-to-beat intervals that differ by more than 50 ms (pNN50) had high positive correlations with the baseline values; SD of normal beat-to-beat intervals (SDNN) and high frequency (HF) had moderate positive correlations, and low frequency (LF) and LF:HF ratio had low positive correlations. The HR, RMSSD, AVNN, and pNN50 had narrow 95% CIs; however, SDNN, LF, HF, and LF:HF ratio had relatively wider 95% CIs. In contrast, the average-per-night test showed that the HR, RMSSD, SDNN, AVNN, pNN50, LF, and HF had high positive relationships (P<.001), and the LF:HF ratio had a moderate positive relationship (P<.001). The average-per-night test also indicated considerably lower error variances than the 5-minute test for the parameters.

Conclusions: The Oura Ring could accurately measure nocturnal HR and RMSSD in both the 5-minute and average-per-night tests. It provided acceptable nocturnal AVNN, pNN50, HF, and SDNN accuracy in the average-per-night test but not in the 5-minute test. In contrast, the LF and LF:HF ratio of the ring had high error rates in both tests.

Keywords: ECG; Oura smart ring; electrocardiography; heart rate variability; wearable device.

PubMed Disclaimer

Conflict of interest statement

Conflicts of Interest: None declared.

Figures

Figure 1
Figure 1
The electrocardiography analysis steps. ECG: electrocardiography; HRV: heart rate variability.
Figure 2
Figure 2
The peak detection result of a 5-minute time window. ECG: electrocardiography.
Figure 3
Figure 3
The pipeline of the electrocardiography validation method. ECG: electrocardiography; HR: heart rate; RR: respiratory rate.
Figure 4
Figure 4
The scatter plots and regression analysis of the nocturnal heart rate and heart rate variability parameters collected from the Oura Ring and Shimmer electrocardiography in 5-minute segments. The regression lines and ideal lines are indicated in red and black, respectively. AVNN: average of all normal-to-normal intervals; ECG: electrocardiography; HF: high frequency; HR: heart rate; LF: low frequency; pNN50: percentage of successive beat-to-beat intervals that differ by more than 50 ms; PPG: photoplethysmography; RMSSD: root mean square of successive differences; SDNN: SD of beat-to-beat intervals.
Figure 5
Figure 5
The Bland–Altman plots of the nocturnal heart rate and heart rate variability parameters in 5-minute segments obtained by the Oura Ring and Shimmer electrocardiography. AVNN: average of all normal-to-normal intervals; HF: high frequency; HR: heart rate; LF: low frequency; pNN50: percentage of successive beat-to-beat intervals that differ by more than 50 ms; PPG: photoplethysmography; RMSSD: root mean square of successive differences; SDNN: SD of beat-to-beat intervals.
Figure 6
Figure 6
The nocturnal heart rate and heart rate variability parameters of one participant during a night sleep event. The data are extracted from the Oura Ring (in red) and Shimmer electrocardiography (in green). AVNN: average of all normal-to-normal intervals; ECG: electrocardiography; HF: high frequency; HR: heart rate; LF: low frequency; pNN50: percentage of successive beat-to-beat intervals that differ by more than 50 ms; PPG: photoplethysmography; RMSSD: root mean square of successive differences; SDNN: SD of beat-to-beat intervals.
Figure 7
Figure 7
The scatter plots and regression analysis of the average heart rate and heart rate variability parameters (which are collected from the Oura Ring and Shimmer electrocardiography) per night sleep time. The regression lines and ideal lines are indicated in red and black, respectively. AVNN: average of all normal-to-normal intervals; ECG: electrocardiography; HF: high frequency; HR: heart rate; LF: low frequency; pNN50: percentage of successive beat-to-beat intervals that differ by more than 50 ms; PPG: photoplethysmography; RMSSD: root mean square of successive differences; SDNN: SD of beat-to-beat intervals.
Figure 8
Figure 8
The Bland–Altman plots of the average heart rate and heart rate variability parameters (which are obtained by the Oura Ring and Shimmer electrocardiography) per night sleep time. AVNN: average of all normal-to-normal intervals; HF: high frequency; HR: heart rate; LF: low frequency; pNN50: percentage of successive beat-to-beat intervals that differ by more than 50 ms; PPG: photoplethysmography; RMSSD: root mean square of successive differences; SDNN: SD of beat-to-beat intervals.

References

    1. McGaughey J, Alderdice F, Fowler R, Kapila A, Mayhew A, Moutray M. Outreach and Early Warning Systems (EWS) for the prevention of intensive care admission and death of critically ill adult patients on general hospital wards. Cochrane Database Syst Rev. 2007 Jul 18;(3):CD005529. doi: 10.1002/14651858.CD005529.pub2. - DOI - PubMed
    1. Anzanpour A, Azimi I, Gotzinger M, Rahmani AM, TaheriNejad N, Liljeberg P, Jantsch A, Dutt N. Self-awareness in remote health monitoring systems using wearable electronics. Proceedings of the Design, Automation & Test in Europe Conference & Exhibition (DATE); Design, Automation & Test in Europe Conference & Exhibition (DATE); Mar 27-31, 2017; Lausanne, Switzerland. 2017. - DOI
    1. McCraty R, Shaffer F. Heart rate variability: new perspectives on physiological mechanisms, assessment of self-regulatory capacity, and health risk. Glob Adv Health Med. 2015 Jan;4(1):46–61. doi: 10.7453/gahmj.2014.073. https://journals.sagepub.com/doi/10.7453/gahmj.2014.073?url_ver=Z39.88-2... gahmj.2014.073 - DOI - DOI - PMC - PubMed
    1. Shaffer F, Ginsberg JP. An overview of heart rate variability metrics and norms. Front Public Health. 2017 Sep 28;5:258. doi: 10.3389/fpubh.2017.00258. doi: 10.3389/fpubh.2017.00258. - DOI - DOI - PMC - PubMed
    1. Song T, Qu XF, Zhang YT, Cao W, Han BH, Li Y, Piao JY, Yin LL, Da Cheng H. Usefulness of the heart-rate variability complex for predicting cardiac mortality after acute myocardial infarction. BMC Cardiovasc Disord. 2014 May 01;14:59. doi: 10.1186/1471-2261-14-59. https://bmccardiovascdisord.biomedcentral.com/articles/10.1186/1471-2261... 1471-2261-14-59 - DOI - DOI - PMC - PubMed

Publication types