Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2022 Jan 21:11:e72629.
doi: 10.7554/eLife.72629.

Nanofluidic chips for cryo-EM structure determination from picoliter sample volumes

Affiliations

Nanofluidic chips for cryo-EM structure determination from picoliter sample volumes

Stefan T Huber et al. Elife. .

Abstract

Cryogenic electron microscopy has become an essential tool for structure determination of biological macromolecules. In practice, the difficulty to reliably prepare samples with uniform ice thickness still represents a barrier for routine high-resolution imaging and limits the current throughput of the technique. We show that a nanofluidic sample support with well-defined geometry can be used to prepare cryo-EM specimens with reproducible ice thickness from picoliter sample volumes. The sample solution is contained in electron-transparent nanochannels that provide uniform thickness gradients without further optimisation and eliminate the potentially destructive air-water interface. We demonstrate the possibility to perform high-resolution structure determination with three standard protein specimens. Nanofabricated sample supports bear potential to automate the cryo-EM workflow, and to explore new frontiers for cryo-EM applications such as time-resolved imaging and high-throughput screening.

Keywords: MEMS; cryo-EM; molecular biophysics; nanofluidics; none; specimen preparation; structural biology.

PubMed Disclaimer

Conflict of interest statement

SH, FW, WE, AJ No competing interests declared, ES, RH ES and RH are inventors on a provisional patent application by SmartTip B.V. (NL2026627) for the use of microcantilevers with nanofluidic sample cells

Figures

Figure 1.
Figure 1.. A nanofluidic sample support for cryo-EM.
(a) Schematic representation of the MEMS-based cryoChip. The 2 × 2 mm chip base features two microcantilevers for sample filling and air escape. The cantilevers are connected to microchannels leading to a free-standing observation membrane located within an aperture at the chip center (middle inset). A multifurcation connects the microchannels (blue) to an array of five parallel nanochannels (yellow) of approximately 700 nm width and 100 nm height (rightmost inset). The top membrane is smoothly sagging, while the bottom membrane is flat. (b) False-colour scanning electron microscopy (SEM) image of the observation membrane. (c) False-colour SEM image of a single nanochannel with FIB-milled rectangular opening across the observation membrane (d) Ellipsometric determination of SiNx membrane thickness after deposition and etching steps. (e) Relative thickness map of an ice-filled nanochannel from zero-loss imaging. WET = water-equivalent thickness, IT = ice thickness. The thickness profile correlates with a 2D slice of a tomogram acquired of the same area (bottom). (f) Thickness cross-sections of nanochannels from three chips of the same wafer. Thin lines are individual nanochannel cross-sections measured across a whole chip; thick lines represent their mean.
Figure 1—figure supplement 1.
Figure 1—figure supplement 1.. cryoChip fabrication.
Fabrication schematic of cryoChips. All steps for nanofabrication of the cryoChip are shown in cross-section. The graphic on the bottom shows the location of microchannels (supply channels for filling) and nanochannels (for imaging) in the schematic.
Figure 1—figure supplement 2.
Figure 1—figure supplement 2.. Specimen mounting.
Preparation of cryoChips for TEM imaging. (a) Mounting in Autogrid cartridges (Thermo Fisher Scientific). A cryoChip is sandwiched between two molybdenum adapter rings with a circular aperture (laser cut from 12.5 µm molybdenum foil) and fixated with the Autogrid clip ring. Chips are mounted with the observation membrane facing the cartridge base and the slanted KOH-etch side walls facing the clip ring (b) Mounting in cartridges for the JEOL JEM3200-FSC microscope. The chip is placed with the observation membrane facing the cartridge base. The default metal spacer ring is placed on top and the assembly is secured by a screw ring.
Figure 2.
Figure 2.. Sample preparation with cryoChips.
(a) Manual filling station with side view and top view microscope, microvolume glass syringe, xyz precision stages for translation of the syringe and a mount for tweezers holding a cryoChip. (b) Sample loading of a cryoChip seen with the top view microscope. One of the cantilevers is approached with a ~40 nL sample droplet. (c) Kymograph of the filling process showing the progression of the sample meniscus over time. The arrow indicates the cantilever and the supply channel. (d) Filling kinetics of cryoChips. The meniscus position and associated time value were recorded at 10 fps with the top-view microscope, revealing a decreasing filling rate with progression of the fluid front (left panel). Each graph represents data from one filling experiment. Graphs from individual experiments are spaced apart by 0.5 s for better visualisation. The squared meniscus position (right panel) displays a linear relationship with elapsed filling time in agreement with expectation from Washburn kinetics. Solid lines are non-linear fits to the data. Slopes (106 μm2/s) and their standard deviations from bootstrapping statistics are shown.
Figure 2—figure supplement 1.
Figure 2—figure supplement 1.. Filling kymographs.
Kymographs showing liquid entry and exit through the supply- and exit microchannels. The liquid takes around one second to flow from the cantilever inside the whole entry channel. Due to the small cross-section of the nanochannels in the observation membrane, liquid flows out of the exit channel very slowly.
Figure 3.
Figure 3.. Automated data acquisition with cryoChips.
(a) TEM image of the observation membrane with five nanochannels. The checkerboard structure is a stiffening layer made from silicon oxide that divides the nanochannels in observation windows. (b) Close-up of inset in (a). (c) Medium magnification image of an observation square with an overlay of the SerialEM multishot acquisition scheme containing six acquisition points along the nanochannel. Each acquisition point is centered on a polygon outlining the limits of three imaging positions achieved by beam-image shift, leading to a total of 18 images per observation square.
Figure 4.
Figure 4.. 3D reconstruction of ApoFtn, T20S proteasome and TMV.
(a) Raw micrographs, (b) 2D class averages, (c) 3D cryo-EM maps and (d) close-up densities displaying discernible details for the three test specimens. Atomic models were overlaid as visual aid: ApoFtn pdb:2×17 (Kasyutich et al., 2010), TMV pdb:4udv (Fromm et al., 2015), T20S pdb:3j9i (Li et al., 2013) (e) Percentage of images in the dataset classified as vitreous (v) and icy (i). (f) Fourier Shell Correlation (FSC) curves with 0.143 cutoff. (g) ResLog plots for reconstructions. B-factors were estimated as two over the fitted slope. Colour coding is equivalent for panels (c–g). Micrographs in (a) were taken at a defocus of –1.2 µm for ApoFtn, –1.8 µm for TMV and –2.8 µm for T20S. Scale bars are equivalent for images in (a) and class averages in (b).
Figure 4—figure supplement 1.
Figure 4—figure supplement 1.. Vitrification efficiency.
Vitrification efficiency analysis. Representative images with vitreous ice (left panels) and crystalline ice (right panels) from the three test specimens. CTF fits to the power spectra are also shown. Thon rings from SiNx can be used for coma-free alignment. Scale bar is 100 nm.
Figure 4—figure supplement 2.
Figure 4—figure supplement 2.. COMSOL simulations.
Simulation of plunge freezing of a simplified model of cryoChips in COMSOL 5.4. (a) Temperature profiles over 50 µm observation membrane from 1-10 µs (pink), 1–10 ms (orange) and 10–100 ms (blue). Dashed line shows glass-transition temperature of water at 136 K. (b) 2D model of cryoChips with 2 mm x 100 µm silicon base and 50 µm x 20 nm SiNx observation membrane used in simulations.
Figure 4—figure supplement 3.
Figure 4—figure supplement 3.. Ice formation.
Analysis of icy micrographs from ApoFtn dataset. (a) Micrograph (top) and power spectrum (bottom) for three examples with crystalline ice. Solid rings indicate expected Bragg spacings diffraction spots for hexagonal ice (white) and cubic ice (red) (derived from Kumai, 2017). (b) Power spectra for frame subsets from the first example in (a) indicate no visible intensity changes of ice reflections throughout the exposure, suggestive of non-vitreous sections in the nanochannels and not surface ice contamination. (c) Maximum intensity projection of all power spectra from the entire ApoFtn dataset. (d) Rotational average of (c) shows four discernible diffraction peaks, including the (100) reflection (3.90 Å) unique for hexagonal ice (blue arrow).
Figure 4—figure supplement 4.
Figure 4—figure supplement 4.. ApoFtn processing.
Single-particle analysis workflow for P. furiosus apoferritin (ApoFtn).
Figure 4—figure supplement 5.
Figure 4—figure supplement 5.. TMV processing.
Single-particle analysis workflow for tobacco mosaic virus (TMV).
Figure 4—figure supplement 6.
Figure 4—figure supplement 6.. T20S processing.
Single-particle analysis workflow for T. acidophilum T20S proteasome (T20S).
Figure 4—figure supplement 7.
Figure 4—figure supplement 7.. B-factor analysis.
Comparison of cryoChip datasets with reference datasets on conventional holey carbon (QF). (a) B-factor plots for ApoFtn, T20S and TMV from cryoChip and holey carbon (T20S and ApoFtn) datasets obtained using the same acquisition settings and processing workflows. (b) Motion trajectories for ApoFtn datasets from cryoChip and holey carbon grids. The overall specimen movement for cryoChips is higher, but initial movement at 0-5 e--2 is on the order of 0.5 pixel/frame. (c) Trajectories from (b). Fourier Shell Correlation (FSC) curves for 3D reconstructions of (d) ApoFtn and (e) T20S from holey carbon reference datasets. datasets.
Figure 5.
Figure 5.. Comparison of beam-induced motion for cryoChip and reference datasets from conventional sample preparation.
(a) Whole-frame motion trajectories of dose-fractionated movies during exposure with 63 e-2 from three different cryoChips. (b) Trajectories coloured by chip or by imaging position in a repeating three-shot pattern. (c) Whole-frame motion trajectories of reference datasets. (d) Root-mean-square deviation (RMSD) of trajectories over electron exposure between the cryoChip dataset and reference datasets. The inset shows the first part of the exposure most critical for reconstructing high-resolution details. Scale bar in (a) is 100 nm. First five e-2 in trajectories are coloured in black.
Figure 5—figure supplement 1.
Figure 5—figure supplement 1.. Motion analysis.
Beam-induced motion of ApoFtn dataset in cryoChips. Motion trajectories separated by chip and colour-coded by acquisition location reveal preferred directions of movement specific for each shot position in the nanochannels. Pictograms show the direction of nanochannels in the acquisition movies. Scale bar: 100 nm.
Figure 6.
Figure 6.. Tomographic reconstruction of a cryoChip nanochannel filled with ApoFtn.
(a) Overview image of tomogram area of a nanochannel segment filled with ApoFtn. (b) Slice through the reconstructed tomogram and yz cross-section across the nanochannel indicated by lines. (c) Segmented reconstruction showing the nanochannel membrane (bottom: pink, top: yellow) and ApoFtn in the channel lumen (blue). The thickness of the membrane is affected by the missing wedge and not to scale. All scale bars: 100 nm.
Figure 6—figure supplement 1.
Figure 6—figure supplement 1.. TMV and T20S cryoChip tomograms.
Tomograms of cryoChips with TMV and T20S. (a) Schematic illustration of volume determination and particle counting for a part of the ApoFtn tomogram. Segmented ApoFtn particles are shown as spheres. Low (top) and medium (center) magnification micrographs and segmented tomograms of nanochannels filled with (b) TMV (EMD-12914) and (c) T20S proteasome (EMD-12917). The cryoChip for T20S has a clover-patterned design of nanochannel and circular observation windows.
Figure 6—figure supplement 2.
Figure 6—figure supplement 2.. SiN Fourier spectra.
Spectral analysis of SiN. Real space image, Fourier amplitude spectrum and detail of Fourier amplitude spectrum for (a) 19 nm silicon nitride membrane adjacent to a nanochannel (b) ApoFtn in vitreous ice in a silicon nitride nanochannel. Arrows in the schematics indicate imaging position. The top and bottom membrane of the nanochannel are between 0 and 100 nm apart, leading to interference of their CTFs resulting in nodes (arrows) in their overall Thon ring pattern. (c) Radially averaged power spectra of three micrographs at different defoci show a broad peak of scattering intensity for amorphous SiN at about 3.9 Å. CTF nodes are visible in the nanochannel Fourier spectra (arrows).
Figure 6—figure supplement 3.
Figure 6—figure supplement 3.. Surface electrostatic potential maps.
Electrostatic potential maps for (a) ApoFtn, displayed along fourfold, threefold, and twofold axes and (b) T20S proteasome displayed along the sevenfold and twofold axes. Electrostatic potentials have been computed from PDB ID 2×17 (Kasyutich et al., 2010) and PDB ID 3j9i (Li et al., 2013) using the Adaptive Poisson Boltzmann server (ABPS) (Jurrus et al., 2018).

References

    1. Adams PD, Afonine PV, Bunkóczi G, Chen VB, Davis IW, Echols N, Headd JJ, Hung L-W, Kapral GJ, Grosse-Kunstleve RW, McCoy AJ, Moriarty NW, Oeffner R, Read RJ, Richardson DC, Richardson JS, Terwilliger TC, Zwart PH. PHENIX: a comprehensive Python-based system for macromolecular structure solution. Acta Crystallographica. Section D, Biological Crystallography. 2010;66:213–221. doi: 10.1107/S0907444909052925. - DOI - PMC - PubMed
    1. Adrian M, Dubochet J, Lepault J, McDowall AW. Cryo-electron microscopy of viruses. Nature. 1984;308:32–36. doi: 10.1038/308032a0. - DOI - PubMed
    1. Angell CA. Liquid fragility and the glass transition in water and aqueous solutions. Chemical Reviews. 2002;102:2627–2650. doi: 10.1021/cr000689q. - DOI - PubMed
    1. Armstrong M, Han BG, Gomez S, Turner J, Fletcher DA, Glaeser RM. Microscale Fluid Behavior during Cryo-EM Sample Blotting. Biophysical Journal. 2020;118:708–719. doi: 10.1016/j.bpj.2019.12.017. - DOI - PMC - PubMed
    1. Arnold SA, Albiez S, Bieri A, Syntychaki A, Adaixo R, McLeod RA, Goldie KN, Stahlberg H, Braun T. Blotting-free and lossless cryo-electron microscopy grid preparation from nanoliter-sized protein samples and single-cell extracts. Journal of Structural Biology. 2017;197:220–226. doi: 10.1016/j.jsb.2016.11.002. - DOI - PubMed

Publication types

MeSH terms