. 2022 Jun;37(6):1130-1135.
doi: 10.1016/j.arth.2022.01.093. Epub 2022 Feb 5.

Unexpected Wear of a Uniquely Designed Moderately Cross-Linked Polyethylene in Total Hip Arthroplasty

Affiliations

Unexpected Wear of a Uniquely Designed Moderately Cross-Linked Polyethylene in Total Hip Arthroplasty

Ramakanth R Yakkanti et al. J Arthroplasty. 2022 Jun.

Abstract

Background: A uniquely designed, non-heat-treated moderately cross-linked acetabular polyethylene liner used in total hip arthroplasty (THA) demonstrated excessive wear during routine follow-up, prompting an evaluation of the linear wear rate.

Methods: All THAs were performed by the senior author. The study group included 38 THAs using the uniquely designed polyethylene in question, compared to a control group of 21 THAs using another moderately cross-linked polyethylene with good 10-year outcomes. Two-dimensional linear head penetration wear measurements were obtained using the Martell Hip Analysis Suite, and retrieval analysis was performed on two liners.

Results: The study group had a significantly higher average penetration rate of 0.089 mm/y than the control group average rate of 0.047 mm/y (P = .04). Forty-five percent of the study group had a wear rate above the osteolysis threshold (0.1 mm/y), compared to 24% in the control group. Macroscopic analysis of two retrieved liners validated the radiographic findings.

Conclusion: The data suggest unexpectedly higher wear rates for a moderately cross-linked polyethylene design, with nearly half of the study group at risk for osteolysis. Further registry or database analyses of this particular moderately cross-linked polyethylene are warranted.

Keywords: acetabular liner; excessive wear; osteolysis threshold; polyethylene wear; total hip arthroplasty.